Z-0130-2014Class II
Genesis(r) II, Non-Porous Tibial Base, Size 5, Left, Qty: (1), Ref 71420168
Smith & Nephew, Inc. / initiated 2013-10-07 / Terminated / root cause: manufacturing process
Smith & Nephew Inc began a recall of GENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 5, LEFT, QTY: (1), REF 71420168. Product Usage: orthopaedic on . FDA classified it as Class II and gives the reason as "Swapped product. Non-porous tibial base size 5 L was mis-packaged as size 6 R and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0130-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- GENESIS(R) II, NON-POROUS TIBIAL BASE, SIZE 5, LEFT, QTY: (1), REF 71420168. Product Usage: orthopaedic
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Swapped product. Non-porous tibial base size 5 L was mis-packaged as size 6 R and vice versa.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Smith & Nephew sent an Urgent - Product Recall 1st Notification letter to all affected sales staff via email and FedEx on October 7, 2013. The hospital was notified by sales rep on October 9, 2013 and the notification letter was sent to the hospital via Fed Ex on October 11, 2013. The letter identified the affected product, problem and actions to be taken. Customers were instructed to inspect inventory, discontinue use, quarantine affected products immediately and return to Smith & Nephew. Customers were asked to complete the attached Inventory Return Certification Form and fax to 901-566-7975.
- Quantity in commerce
- 4 units
- Distribution
- USA Nationwide Distribution in the states of: AZ, NC, IA, PA, OH, MD, and TN.
- Product codes and lots
- Batch No. 13GM11272
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0130-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0131-2014 | Class II | Terminated | Voluntary: Firm initiated |