Z-0187-2014Class II

Safire Duo Ablation Catheter MediGuide Enabled, REF A700240 and A700238, Rx only, Sterile EO. Intended for...

Abbott Medical / initiated 2013-10-11 / Terminated

St. Jude Medical began a recall of Safire Duo Ablation Catheter MediGuide Enabled, REF A700240 and A700238, Rx only, Sterile EO. Intended for creating endocardial lesions during cardiac ablation... on . FDA classified it as Class II and gives the reason as "St. Jude Medical Cardiovascular and Ablation Technologies Division is performing a voluntary notification regarding selected batched of Safire Duo Ablation Catheter MediGuide Enabled. These units have a cable connector...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0187-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Safire Duo Ablation Catheter MediGuide Enabled, REF A700240 and A700238, Rx only, Sterile EO. Intended for creating endocardial lesions during cardiac ablation procedures (mapping, stimulation and ablation) for the treatment of typical atrial flutter.
Product code
OAD Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter
Reason for recall
St. Jude Medical Cardiovascular and Ablation Technologies Division is performing a voluntary notification regarding selected batched of Safire Duo Ablation Catheter MediGuide Enabled. These units have a cable connector configuration that is different than intended.
FDA root cause
Process change control
Action
Consignees were sent a St. Jude Medical "Urgent Medical Device Notice" dated 11 October 2013. The letter was addressed to Cath lab manger / Risk Manger. The letter described the problem and the product involved in the recall. The letter stated that it is acceptable to continue using these devices as long as the product is used in accordance with the IFU.
Quantity in commerce
46
Distribution
Nationwide Distribution including the states of KY, OH, TX, MA, and NC.
Product codes and lots
Catalog Number A700238, Batch Number 4113243, 4152911. Catalog Number A700240, Batch Number 4113242, 4124985.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0187-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0187-2014Class IITerminatedVoluntary: Firm initiated