Z-0802-2014Class II

Civco Belly Board, Reusable non-sterile carbon fiber belly board with cushion, REF 125012, packaged 1 per box

CIVCO Medical Instruments Co., Inc. / initiated 2013-10-25 / Terminated / root cause: design

Civco Medical Instruments Inc began a recall of Civco Belly Board, Reusable non-sterile carbon fiber belly board with cushion, REF 125012, packaged 1 per box. Product Usage: This device is intended to reduce... on . FDA classified it as Class II and gives the reason as "Scales adhered to the product may be misaligned from the zero reference such that one side will not match the other". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0802-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CIVCO Medical Instruments Co., Inc.
Product
Civco Belly Board, Reusable non-sterile carbon fiber belly board with cushion, REF 125012, packaged 1 per box. Product Usage: This device is intended to reduce the irradiated small bowel volume of gynecological patients undergoing treatment in the pelvic region.
Product code
IYE Accelerator, Linear, Medical
Reason for recall
Scales adhered to the product may be misaligned from the zero reference such that one side will not match the other.
Root cause tag
design
FDA root cause
Device Design
Action
Civco Medical Instruments Inc sent a FIELD SAFETY NOTIFICATION letter dated October 28, 2013 via UPS to their customers. The notification identified the affected device, problem, and field safety corrective action. The letter describes how customers can check to see whether the scales on their device are out of alignment. The letter states CIVCO will rework the system to correct the difference. For questions contact sales representative at CIVCO Medical Solutions, Radiation Oncology at 800-842-8688 or 712-737-8688.
Quantity in commerce
138 belly boards
Distribution
Worldwide Distribution - USA Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, and WV. Foreign distribution was made to Brazil, Canada, Chile, Czech Republic, France, Germany, Iraq, Israel, Italy, Japan, Malaysia, Saudi Arabia, Slovenia, South Korea, Spain, Turkey, and United Arab Emirates.
Product codes and lots
Lot #'s: M441670, M469970, M469980, M469990, M484910, M484920, M516200, M516210, M516220, M533340, M533350, M533360, M533370, M533380, M533970, M533980, M533990, M534000, M544650, M544660, M544670, M550150, M551490, M554940, M554950, M556540, M556550, M559560, M560740, M575850, M578310, M581730, M591090, M591100, M591110, M591120, M591130, and M599570

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0802-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0802-2014Class IITerminatedVoluntary: Firm initiated
Z-0803-2014Class IITerminatedVoluntary: Firm initiated