Civco Belly Board MRI, Reusable non-sterile glass fiber belly board with cushion, REF 126000, packaged 1 per...
Civco Medical Instruments Inc began a recall of Civco Belly Board MRI, Reusable non-sterile glass fiber belly board with cushion, REF 126000, packaged 1 per box. Product Usage: This device is intended to... on . FDA classified it as Class II and gives the reason as "Scales adhered to the product may be misaligned from the zero reference such that one side will not match the other". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0803-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CIVCO Medical Instruments Co., Inc.
- Product
- Civco Belly Board MRI, Reusable non-sterile glass fiber belly board with cushion, REF 126000, packaged 1 per box. Product Usage: This device is intended to reduce the irradiated small bowel volume of gynecological patients undergoing treatment in the pelvic region.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Scales adhered to the product may be misaligned from the zero reference such that one side will not match the other.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Civco Medical Instruments Inc sent a FIELD SAFETY NOTIFICATION letter dated October 28, 2013 via UPS to their customers. The notification identified the affected device, problem, and field safety corrective action. The letter describes how customers can check to see whether the scales on their device are out of alignment. The letter states CIVCO will rework the system to correct the difference. For questions contact sales representative at CIVCO Medical Solutions, Radiation Oncology at 800-842-8688 or 712-737-8688.
- Quantity in commerce
- 2 belly board
- Distribution
- Worldwide Distribution - USA Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, and WV. Foreign distribution was made to Brazil, Canada, Chile, Czech Republic, France, Germany, Iraq, Israel, Italy, Japan, Malaysia, Saudi Arabia, Slovenia, South Korea, Spain, Turkey, and United Arab Emirates.
- Product codes and lots
- Lot#'s: M551490 and M550150
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0802-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0803-2014 | Class II | Terminated | Voluntary: Firm initiated |