US Safire Duo Ablation Catheter MediGuide Enabled, REF A700240 and Cool Path Duo Ablation Catheter, MediGuide...

Abbott Medical / initiated 2013-10-11 / Terminated / root cause: labeling and UDI

St. Jude Medical began a recall of US Safire Duo Ablation Catheter MediGuide Enabled, REF A700240 and Cool Path Duo Ablation Catheter, MediGuide Enabled REF A700244, Rx only, Sterile EO. The... on . FDA classified it as Class II and gives the reason as "St. Jude Medical Cardiovascular and Ablation Technologies Division is performing a voluntary notification regarding selected batches of Coolpath Duo Ablation Catheter MediGuide Enabled products and Safire Duo Ablation...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0174-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
US Safire Duo Ablation Catheter MediGuide Enabled, REF A700240 and Cool Path Duo Ablation Catheter, MediGuide Enabled REF A700244, Rx only, Sterile EO. The catheters are intended for use with a compatible external irrigation pump and the IBI-15500T9-CP RF Generator at a maximum of 50 Watts. MediGuide Enabled Ablation catheters are used with the MediGuide Technology to enable real-time tip positioning and navigation and in conjunction the the EnSite Velocity System. The MediGuide Technology is indicated for use as an adjunct to fluoroscopy.
Product code
OAD Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter
Reason for recall
St. Jude Medical Cardiovascular and Ablation Technologies Division is performing a voluntary notification regarding selected batches of Coolpath Duo Ablation Catheter MediGuide Enabled products and Safire Duo Ablation Catheter MediGuide Enabled. These units were inadvertently shipped with the incorrect version of the Instructions for Use (IFU).
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
The firm, St. Jude Medical (SJM), sent a "Urgent Medical Device Notice" dated 11 October 2013 to its consignees. The letter described the problem, the product involved in the recall and the actions to be taken. The consignees were instructed to discard the IFU received with the batches and replace it with the IFU received in conjunction with the notification letter. SJM has determined that it is acceptable to continue using these devices as long as the product is used in accordance with the IFU. If you have any questions, contact Vice President, Quality, at 651-756-5402.
Quantity in commerce
59
Distribution
US distribution: KS, MA, UT,TX, KY, OH.
Product codes and lots
Catalog Number A700244 Batch Number 4113241. Catalog Number A700240 Batch Number 4113242.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0174-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0174-2014Class IITerminatedVoluntary: Firm initiated