Z-2609-2016Class II

1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part...

Baxter Healthcare Corporation / initiated 2013-10-07 / Terminated / root cause: labeling and UDI

Baxter Healthcare Corp. began a recall of 1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal... on . FDA classified it as Class II and gives the reason as "In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2609-2016
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
1) The HomeChoice automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. 2) The HomeChoice PRO automated peritoneal dialysis system. Dataplate located on the device is labeled in part: Baxter. The HomeChoice/HomeChoice Pro Automated Personal Cycler peritoneal dialysis system is intended for automatic control of dialysate solution exchange in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
Product code
FKX System, Peritoneal, Automatic Delivery
Reason for recall
In systems with version 10.4 software Initial Drain logic, the device will attempt to drain the patient to empty at the beginning of therapy (initial drain). The operator can stop, but cannot bypass, the active Initial Drain. This is to mitigate against Unintended Increased Intraperitoneal Volume (IIVP). This can cause serious problems in patients with unrelated a co-morbid condition of ascites.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
An IMPORTANT PRODUCT INFORMATION letter dated October 7, 2013, was sent to all affected customers. The letter included instructions for clinicians to: 1) be aware that HomeChoice will attempt to drain the patient to empty at the beginning of therapy (Initial Drain); 2) review HomeChoice instructions for use for additional details on HomeChoice 10.4 Initial Drain Logic; 3) acknowledge your receipt of the notification by completing the attached Customer Reply Form and returning by fax to either 224-270-5457 or scanning and e-mailing it to [email protected]; 3) if the recipient of the letter is a dealer, wholesaler, or distributor/reseller that distributed any product to other facilities, to forward the communication to all customers who may have received these products. Clinical questions should be addressed to Baxter's Renal Division Clinical Helpline at 888-736-2543, option 2 (Monday - Friday, 8:00 AM - 5:00PM CST). General questions about this recall should be addressed to The Center for One Baxter at 800-422-9837 (Monday - Friday, 8:00 AM - 5:00 PM CST).
Quantity in commerce
*** 1) Product Codes 5C4471 and 5C4471R: Approximately 48,600 units; *** 2) Product Codes 5C8310 and 5C8310R: Approxiamtely 16,990 units
Distribution
Worldwide distribution. US nationwide including Puerto Rico; Australia, Canada, New Zealand, and Japan
Product codes and lots
*** 1) HomeChoice automated peritoneal dialysis system:  Product Codes: 5C4471 and 5C4471R;  Serial Numbers: All serial numbers;   *** 2) HomeChoice Pro automated peritoneal dialysis system:  Product Codes: 5C8310 and 5C8310R;  Serial Numbers: All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2609-2016
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2609-2016Class IITerminatedVoluntary: Firm initiated