Applied Medical's Inzii 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is...
Applied Medical Resources Corp began a recall of Applied Medical's Inzii 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared... on . FDA classified it as Class II and gives the reason as "Applied Medical is conducting a voluntary recall of the Inzii 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0183-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Applied Medical Resources Corp
- Product
- Applied Medical's Inzii 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared to market via 510(k) filing K060051. The retrieval bag is a disposable specimen retrieval bag intended for use as receptacle for the collection of tissue, organs, and calculi during laparoscopic procedures. Applieds CD004 has a shelf life of three years. It is provided sterile and packaged in a heat sealed Tyvek pouch.
- Product code
- GCJ Laparoscope, General & Plastic Surgery
- Reason for recall
- Applied Medical is conducting a voluntary recall of the Inzii 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen
- Root cause tag
- packaging
- FDA root cause
- Package design/selection
- Action
- Customer notifications of the recall began the week of 10/21/2013 via UPS overnight with letters requesting that all affected products be returned to Applied. The notification letter is dated October 21, 2013 and titled "URGENT: MEDICAL DEVICE RECALL". The notification was accompanied by a Recall Notification Confirmation Form. The letter informed customers of the recall by providing the product description with codes (Model and Lot Numbers). reason for the recall, actions to take and contact information.
- Quantity in commerce
- 29,506 units
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Lebanon, Malaysia, Mexico, New Zealand, Netherlands, Norway, Peru, Poland, Portugal, Saudi Arabia, Singapore, Spain, South Africa, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
- Product codes and lots
- Model Number CD004. Lot Nubmers: 1161268, 1161269, 1164934, 1168358, 1168361, 1169317, 1169318, 1170694, 1170700, 1170701, 1171214, 1172482, 1172664, 1174287, 1174793, 1174857, 1175123, 1175280, 1175472, 1176555, 1177023, 1177871, 1179069, 1179392, 1179460, 1179775, 1179890, 1180291, 1180411, 1180576, 1181647, 1181648, 1181649, 1181939, 1182760, 1182934, 1183225, 1183823, 1183957, 1184273, 1184728, 1184915, 1184916, 1187779, 1188817, 1189117, 1189957, 1190294, 1190358, 1191208, 1192057, 1192969, 1193196, 1193661, 1194010, 1194666, 1194807, 1195518, 1195727, 1195997, 1197182, 1197404, 1197832, 1198055, 1198968, 1199899, 1200658, 1200828, 1200829, 1200830, 1201739, 1201740, 1201741, 1203154, 1204054.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0183-2014 | Class II | Terminated | Voluntary: Firm initiated |