Fresenius 2008 Series Hemodialysis Machines: Models: 2008H, 2008K, 2008K2, 2008k@Home, 2008T The 2008H...

Fresenius Medical Care Renal Therapies Group, LLC / initiated 2013-11-12 / Terminated

Fresenius Medical Care Holdings, Inc. began a recall of Fresenius 2008 Series Hemodialysis Machines: Models: 2008H, 2008K, 2008K2, 2008k@Home, 2008T The 2008H Hemodialysis machine is indicated for acute and chronic... on . FDA classified it as Class II and gives the reason as "2008 Series Hemodialysis Machines complaints of saline bags inappropriately filling during recirculation and setup". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0629-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Product
Fresenius 2008 Series Hemodialysis Machines: Models: 2008H, 2008K, 2008K2, 2008k@Home, 2008T The 2008H Hemodialysis machine is indicated for acute and chronic dialysis therapy. Hemodialysis is indicated for patients with acute or chronic renal failure, when conservative therapy is judged to be inadequate. The dialysate delivery machine is used for hemodialysis in hospital, dialysis centers, and at home. The 2008K, 20081(2 and 2008T hemodialysis machines are designed to perform hemodialysis in hospitals and dialysis clinics. They can be used for patients suffering chronic or acute renal failure. The 2008K, 2008K2 and 2008T are designed to provide hemodialysis treatment by controlling and monitoring both the dialysate and extracorporeal blood circuits. The Fresenius 2008K~home is indicated for acute and chronic dialysis therapy in an acute or chronic facility. The 2008K~home is also indicated for hemodialysis in the home and must be observed by a trained and qualified person as prescribed by their physician.
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
2008 Series Hemodialysis Machines complaints of saline bags inappropriately filling during recirculation and setup
FDA root cause
Under Investigation by firm
Action
Fresenius Medical issued a Product Notification letter dated November 12, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers who encounter a saline bag that is inappropriately filling, should discontinue the setup and remove the machine from service until the drain line and drain-associated hydraulic components can be checed by a qualified service technician. Customers were also instructed to ensure that the hemodialysis machine is installed and operated in compliance with the Machine Specification section of their operator's manual, including the specification located in the Water/Drain Section. Customers with questions were instructed to call 1-800-227-2572. For questions regarding this recall call 800-662-1237.
Quantity in commerce
111,504 units US and 3,638 Canada and Mexico-
Distribution
Worldwide Distribution - USA (nationwide) and Internationally Mexico and Canada.
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0629-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0629-2014Class IITerminatedVoluntary: Firm initiated