Sarns" Venous Return Cannulae, 20 Fr with 1/4" flare, 14.5" (37 cm) long. Indicated for venous drainage...
Terumo Cardiovascular Systems Corporation began a recall of Sarns" Venous Return Cannulae, 20 Fr with 1/4" flare, 14.5" (37 cm) long. Indicated for venous drainage during cardiopulmonary bypass surgery for dial... on . FDA classified it as Class II and gives the reason as "During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0196-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- Sarns" Venous Return Cannulae, 20 Fr with 1/4" flare, 14.5" (37 cm) long. Indicated for venous drainage during cardiopulmonary bypass surgery for dial cannulation of the superior and inferior vena cava.
- Product code
- DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
- Reason for recall
- During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow Aortic Cannulae and Sarns" Venous Return Cannulae.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- On 10/23/13, Terumo Cardiovascular Systems sent an URGENT MEDICAL DEVICE RECALL LETTER to their consignees. The letter identified the reason for the device removal, identified the affected product and associated potential hazards, and provided recall instructions to their customers regarding affected product return. Questions regarding this recall are directed to CVS Customer Service M-F, 8am-6pm at 1-800-521-2818.
- Quantity in commerce
- 40 units
- Distribution
- Worldwide Distribution-USA including DC and the states of MO,TX,OK, LA, MA, CA, WI, and MI and the country of Canada.
- Product codes and lots
- Catalog Number 9473 Lot Numbers 0689812
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0193-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0194-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0195-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0196-2014 | Class II | Terminated | Voluntary: Firm initiated |