Z-1089-2014Class II

Azure Anterior Cervical Plate System, 2-Level Plate, 36 mm, Implant Grade Titanium and Nitinol, Orthofix...

Orthofix SRL / initiated 2013-10-29 / Terminated / root cause: design

Orthofix, Inc began a recall of Azure Anterior Cervical Plate System, 2-Level Plate, 36 mm, Implant Grade Titanium and Nitinol, Orthofix Inc., RX Only The Azure Anterior Cervical Plate System... on . FDA classified it as Class II and gives the reason as "There is a possibility that the locking mechanism of the Azure Anterior Cervical Plate may fracture during or after implantation". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1089-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
Azure Anterior Cervical Plate System, 2-Level Plate, 36 mm, Implant Grade Titanium and Nitinol, Orthofix Inc., RX Only The Azure Anterior Cervical Plate System is a temporary implant, intended for anterior fixation to the cervical spine from C2 to C7 and is indicated for: A. Degenerative disc disease, B. Spondylolisthesis, C. Trauma, D. Spinal Stenosis, E. Deformities, F. Tumor, G. Pseudoarthrosis, and H. Revision of previous surgery. The Azure Anterior Cervical Plate System is comprised of a variety of non-sterile, single use, titanium alloy and nitinol components. The system is attached to the anterior aspect of the vertebral body by means of screw to to the cervical spine.
Product code
KWQ Appliance, Fixation, Spinal Intervertebral Body
Reason for recall
There is a possibility that the locking mechanism of the Azure Anterior Cervical Plate may fracture during or after implantation.
Root cause tag
design
FDA root cause
Component design/selection
Action
Orthofix sent a Urgent Medical Device Recall Notification dated October 29, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Immediate Action: 1. Immediately cease any further distribution or use of the AZURE" plates. 2. Notify your surgeons and staff customers that Azure" plates cannot be used and are to be returned to Orthofix. at no cost to you, see Instructions below. 3. Please complete the attached reply form indicating that you have received this letter. Product Return Instructions: To return the Azure plates, please follow these Instructions. 1. Check you inventory of Azura Implant Trays.. 2. Complete the attached form with the quantity you are returning. 3. Contact your local Orthofix representative or Orthofix customer service representative at 888--298-5700 for return instructions and shipment information and replacement with a different Orthofix Anterior Cervical Plate System. .
Quantity in commerce
4,172 units
Distribution
Worldwide Distribution: USA (nationwide) and the country of Germany.
Product codes and lots
86-01xx, 86-02xx, 86-03xx, 86-04xx, 86-05xx  All lot numbers manufactured to date

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1089-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1089-2014Class IITerminatedVoluntary: Firm initiated