EPIQ 7 Ultrasound System, Model 795200, 795201, 795202. Intended for diagnostic ultrasound imaging and fluid...
Philips Medical Systems North America Inc. began a recall of EPIQ 7 Ultrasound System, Model 795200, 795201, 795202. Intended for diagnostic ultrasound imaging and fluid flow analysis on . FDA classified it as Class II and gives the reason as "A problem has been detected with Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. Using the International System Date Format may result in calculation errors in Gestational Age Gestational Age...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0368-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- EPIQ 7 Ultrasound System, Model 795200, 795201, 795202. Intended for diagnostic ultrasound imaging and fluid flow analysis.
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- A problem has been detected with Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. Using the International System Date Format may result in calculation errors in Gestational Age Gestational Age (GA) and Established Due Date (EDD).
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Affected customers will be notified via the Urgent-Field Safety Notice (FSN79500286), Philips Model EPIQ 7 Ultrasound system, dated 2013 October. Customers are informed that using the International System Date Format in the Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. may result in calculation errors in Gestational Age and Established Due Date (EDD). Philips will upgrade customer systems to software version 1.1, which corrects the issue, for a free of charge.
- Quantity in commerce
- 306 units total (206 international; 100 in the US).
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of CT, HI, IL, IN, KY, MA, ME, MI, NC, OR, PA, TX, and WA, and the countries of Australia, Austria, Belgium, Chile, Denmark, Finland, France, Germany, India, Ireland, Republic of Korea, Latvia, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, and United Kingdom.
- Product codes and lots
- TRACE ITEM SERIAL NUMBERS in the UNITED STATES: US713B0171, US713B0207, US813B0017, US813B0018, US813B0019, US813B0020, US813B0022, US813B0023, US813B0024, US813B0025, US813B0122, US813B0123, US813B0125, US813B0127, US813B0128, US813B0129, US813B0131, US813B0132, US813B0134, US813B0135, US813B0233, US813B0234, US813B0235, US813B0236, US813B0237, US813B0238, US813B0239, US813B0240, US813B0423, US813B0424, US813B0425, US813B0426, US813B0427, US813B0428, US813B0429, US813B0430, US813B0431, US813B0432, US813B0433, US813B0434, US913B0033, US913B0034, US913B0035, US913B0036, US913B0037, US913B0038, US913B0039, US913B0040, US913B0057, US913B0060, US913B0061, US913B0062, US913B0063, US913B0064, US913B0065, US913B0066, US913B0106, US913B0107, US913B0108, US913B0109, US913B0110, US913B0111, US913B0112, US913B0113, US913B0144, US913B0176, US913B0201, US913B0256, US913B0257, US913B0350, US913B0363, US913B0364, US913B0365, US913B0366, US913B0367, US913B0368, US913B0369, US913B0380, US913B0384, US913B0395, US913B0396, US913B0397, US913B0398, US913B0399, US913B0400, US913B0401, US913B0402, US913B0409, US913B0441, US913B0444, US913B0445, US913B0446, US913B0447, US913B0503, US913B0522, US913B0614, US913B0615, US913B0616, US913B0618, US913B0619, and USO13B0043. TRACE ITEM SERIAL NUMBERS INTERNATIONAL: US513B0224, US513B0295, US513B0387, US513B0560, US613B0299, US713B0085, US713B0086, US713B0118, US713B0119, US713B0120, US713B0122, US713B0123, US713B0126, US713B0149, US713B0171, US713B0172, US713B0192, US713B0193, US713B0194, US713B0195, US713B0204, US713B0205, US713B0206, US713B0208, US713B0324, US713B0325, US713B0358, US713B0359, US713B0437, US713B0439, US713B0440, US713B0532, US713B0537, US713B0573, US713B0574, US713B0575, US713B0576, US713B0577, US713B0578, US713B0579, US713B0580, US713B0581, US713B0582, US713B0583, US713B0584, US713B0586, US713B0587, US713B0588, US713B0589, US713B0590, US713B0592, US713B0593, US713B0594, US713B0595, US813B0015, US813B0016, US813B0032, US813B0037, US813B0038, US813B0039, US813B0121, US813B0151, US813B0152, US813B0153, US813B0154, US813B0155, US813B0156, US813B0157, US813B0158, US813B0159, US813B0160, US813B0161, US813B0162, US813B0163, US813B0164, US813B0183, US813B0184, US813B0185, US813B0186, US813B0187, US813B0188, US813B0189, US813B0190, US813B0191, US813B0192, US813B0193, US813B0194, US813B0195, US813B0196, US813B0232, US813B0396, US813B0397, US813B0398, US813B0399, US813B0400, US813B0401, US813B0402, US813B0403, US813B0404, US813B0405, US813B0406, US813B0407, US813B0408, US813B0409, US813B0410, US813B0411, US813B0468, US813B0469, US813B0470, US813B0471, US813B0472, US813B0473, US813B0474, US813B0475, US813B0476, US813B0477, US813B0478, US813B0479, US813B0480, US913B0026, US913B0027, US913B0028, US913B0029, US913B0030, US913B0031, US913B0032, US913B0041, US913B0051, US913B0059, US913B0085, US913B0086, US913B0101, US913B0102, US913B0103, US913B0104, US913B0105, US913B0130, US913B0132, US913B0133, US913B0134, US913B0135, US913B0137, US913B0138, US913B0139, US913B0140, US913B0141, US913B0142, US913B0143, US913B0177, US913B0178, US913B0179, US913B0185, US913B0202, US913B0216, US913B0217, US913B0218, US913B0219, US913B0220, US913B0221, US913B0222, US913B0223, US913B0232, US913B0252, US913B0253, US913B0254, US913B0255, US913B0259, US913B0260, US913B0299, US913B0300, US913B0301, US913B0302, US913B0303, US913B0316, US913B0317, US913B0318, US913B0319, US913B0332, US913B0333, US913B0336, US913B0337, US913B0338, US913B0339, US913B0340, US913B0349, US913B0359, US913B0360, US913B0385, US913B0408, US913B0458, US913B0459, US913B0486, US913B0487, US913B0488, US913B0489, US913B0490, US913B0499, US913B0504, US913B0505, US913B0506, US913B0565, US913B0566, US913B0567, US913B0568, USO13B0040, and USO13B0041.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0368-2014 | Class II | Terminated | Voluntary: Firm initiated |