Product Brand Name(s): Alere Cholestech LDX Lipid Profile"GLU Test Cassette, Model 10-991. Product Generic...
Alere San Diego, Inc. began a recall of Product Brand Name(s): Alere Cholestech LDX Lipid Profile"GLU Test Cassette, Model 10-991. Product Generic Name: Alere Cholestech LDX Lipid Profile"GLU Test... on . FDA classified it as Class III and gives the reason as "Alere initiated a recall of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 320167, becuase an incorrect lot number was printed on the outer 10 pack kit box label. The incorrect Lot #...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0379-2014
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Product
- Product Brand Name(s): Alere Cholestech LDX Lipid Profile"GLU Test Cassette, Model 10-991. Product Generic Name: Alere Cholestech LDX Lipid Profile"GLU Test Cassette. Description of the product: Lipid. Glucose panel Test cassettes. Product packaging: 10 individually pouched cassettes in a labeled kit box. Intended for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides, and/or glucose in whole blood, serum, and plasma.
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- Alere initiated a recall of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 320167, becuase an incorrect lot number was printed on the outer 10 pack kit box label. The incorrect Lot # number printed was 321067.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- Alere initiated this recall by sending out notification letters to customers via fax, email, or direct mail. The letter titled "URGENT MEDICAL DEVICE NOTICE", dated October 23, 2013 is accompanied by a response form/Verification form. The notification informed customers of the reason for recall, product description with codes, required actions by the customers, and contact information.
- Quantity in commerce
- 3,621 kits
- Distribution
- Worldwide Distribution: USA (nationwide) and Internationally to Canada.
- Product codes and lots
- Model Number: 10-991. Lot #320167. Expiration date, or Expected shelf life: 8/31/14. Expected shelf life: up to 12 months.