3.7mm and 5.0mm Dynamic Locking Screw (DLS)
Synthes USA HQ, Inc. began a recall of 3.7mm and 5.0mm Dynamic Locking Screw (DLS) Product Usage: 3.7mm and 5.0mm Dynamic Locking Screw (DLS) in combination with Synthes Locking Compression Plates... on . FDA classified it as Class II and gives the reason as "All lots of 3.7mm and 5.0mm Dynamic Locking Screw (DLS) were recalled due to complaints of breakages at the distal tip of the pin identified after successful healing and during planned removal of the device". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0390-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- 3.7mm and 5.0mm Dynamic Locking Screw (DLS) Product Usage: 3.7mm and 5.0mm Dynamic Locking Screw (DLS) in combination with Synthes Locking Compression Plates (LCP) are intended for use in long bone fractures, the fixation of osteopenic bone, the fixation of osteotomies, and for the fixation of non-unions and malunions.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- All lots of 3.7mm and 5.0mm Dynamic Locking Screw (DLS) were recalled due to complaints of breakages at the distal tip of the pin identified after successful healing and during planned removal of the device.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- Synthes sent an Urgent Notice: Medical Device Recall letter to Sales Reps and End Users on July 22, 2013. The letter identified the affect product, problem and actions to be taken. For questions call the Recall center at 610-719-5450.
- Quantity in commerce
- 40248
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Canada.
- Product codes and lots
- All lots of parts 09.213.022S - 09.213.070S, and 09.223.032S- 09.223.090S.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0390-2014 | Class II | Terminated | Voluntary: Firm initiated |