510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0336-2014Class II

Vascutek Gelsoft Vascular Graft (peripheral) Labeling reads in part, "***Gelsoft Gelatin Impregnated Knitted...

Terumo Cardiovascular Systems Corporation / initiated 2013-10-23 / Terminated / root cause: labeling and UDI

Terumo Cardiovascular Systems Corporation began a recall of Vascutek Gelsoft Vascular Graft (peripheral) Labeling reads in part, "***Gelsoft Gelatin Impregnated Knitted Vascular... on . FDA classified it as Class II and gives the reason as "Terumo Cardiovascular Systems (Terumo CVS) became aware of a customer complaint that a 15mm x 8mm Gelsoft graft was received which the customer believed was a 15mm x 6mm Gelsoft graft unit. The manufacturer confirmed a...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0336-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Vascutek Gelsoft Vascular Graft (peripheral) Labeling reads in part, "***Gelsoft Gelatin Impregnated Knitted Vascular Prosthesis***STRAIGHT***STERILE***EO***Item No: 631508***Diameter 8mm***Usable Length 15cm***Manufactured By***VASCUTEK Ltd. a TERUMO Company Newmains Avenue Inchinnan, Renfrewshire PA4 9RR Scotland***" The Vascutek Gelsoft Vascular Prosthesis is a gelatin-sealed, woven polyester graft indicated for abdominal and peripheral vascular repair, i.e. replacement or bypass in aneurysmal and occlusive disease of arteries
Product code
DSY Prosthesis, Vascular Graft, Of 6mm And Greater Diameter
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) became aware of a customer complaint that a 15mm x 8mm Gelsoft graft was received which the customer believed was a 15mm x 6mm Gelsoft graft unit. The manufacturer confirmed a batch of product code 631508 was incorrectly labeled.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Terumo CVS called affected consignees (via telephone)on 10/23/2013 of this device removal. Terumo CVS advised users to discontinue use and return all affected grafts in inventory. Terumo will replace or issue credit for returned grafts. Terumo provided consignees with the reason of the removal, potential hazard, and description of affected product. Terumo addressed any questions from the consignees. Consignees with questions were instructed to contact Terumo CVS Customer Service at 800-521-2818. For questions regarding this recall call 734-741-6173.
Quantity in commerce
9
Distribution
US Distribution in the states of: NC and CA.
Product codes and lots
Catalog Number 631508 Serial Numbers: 0001765391, 0001765392, 0001765393, 0001765394, 0001765395, 0001765396, 0001765397, 0001765398, 0001765399

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0336-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0336-2014Class IITerminatedVoluntary: Firm initiated