Z-0823-2014Class II

Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink system software is a...

Siemens Healthcare Diagnostics Inc. / initiated 2013-10-30 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink system software is a network solution provider and multi-system data... on . FDA classified it as Class II and gives the reason as "Under extremely rare circumstances, a patient result that had been previously disabled may be released without proper review. (Disabled Test in CentraLink System or Aptio System May Become Enabled in CentraLink System.)". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0823-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.
Product code
JQP Calculator/Data Processing Module, For Clinical Use
Reason for recall
Under extremely rare circumstances, a patient result that had been previously disabled may be released without proper review. (Disabled Test in CentraLink System or Aptio System May Become Enabled in CentraLink System.)
Root cause tag
software
FDA root cause
Software design
Action
The firm, Siemens, sent an "Urgent Medical Device Correction" letter dated October 23, 2013 to its customers. The firm also sent a follow-up letter dated October 30, 2013. The letter describes the product, problem and actions to be taken. The customers were instructed Until a service visit is scheduled, do the following: Instruct the laboratory staff to recognize this sequence of events and Ensure that whenever the operator disables a method on an Aptio system or a CentraLinksystem and replenishes reagent for that method, the operator then confirms that the method remains disabled in the CentraLink system or disables it again. The operator must allow up to five minutes to verify that the test remains disabled. In addition, the customers were instructed tocomplete and return the FIELD CORRECTION EFFECTIVENESS CHECK response form via fax to Siemens Healthcare Diagnostics Customer Care Center at 800-441-9250, and to retain this letter with your laboratory records, and forward this letter to those who may have received this product. After a service visit is performed and an interim solution is implemented, the laboratory staff will be trained on this update to understand the minor but necessary changes in the workflow. If you have any questions, please contact your Siemens Customer Care Center at 800-441-9250 or your local Siemens representative.
Quantity in commerce
40
Distribution
US Distribution in the states of: MN, NJ, and NV.
Product codes and lots
CentraLink" Data Management System V14.0.4, V14.0.5 or V14.0.8 Catalog Numbers: 10810210, 10811596, 10814296, 10814298, 10814877, 10815474.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0823-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0823-2014Class IITerminatedVoluntary: Firm initiated