Z-0626-2014Class II

HbA1c Flex(R) reagent cartridge (K3105A, Siemens Material Number 10470481) for the Dimension(R) Clinical...

Siemens Healthcare Diagnostics Inc. / initiated 2013-04-22 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of HbA1c Flex(R) reagent cartridge (K3105A, Siemens Material Number 10470481) for the Dimension(R) Clinical Systems. Dimension(R) Hemoglobin A1c (HB1C) Flex(R)... on . FDA classified it as Class II and gives the reason as "Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0626-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
HbA1c Flex(R) reagent cartridge (K3105A, Siemens Material Number 10470481) for the Dimension(R) Clinical Systems. Dimension(R) Hemoglobin A1c (HB1C) Flex(R) reagent Cartridge on the Dimension(R) clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long term glucose control in individuals with diabetes mellitus.
Product code
LCP Assay, Glycosylated Hemoglobin
Reason for recall
Positive bias on the lots listed compared to alternate methods and a higher frequency of "Above Assay Range" flags with the lots listed.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Siemens sent an Urgent Medical Device Recall letter in April 2013 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately discard any remaining inventory of the affected product. Siemens will replace any unused inventory at no charge. Customers were asked to complete the attached form and fax it to 302-631-8467 to indicate they have received the information. Customers were asked to retain the letter with their laboratory records and forward the letter to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Technical Solutions Center or their local Siemens technical support representative. For questions regarding this recall call 1-800-441-9250.
Quantity in commerce
11006 Total
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Canada and Mexico.
Product codes and lots
Catalog number K3105A (Siemens Material Number 10470481) - Lot numbers 12072AA, exp. 2/9/13: 12086AA, exp. 2/23/13:12107AA, exp. 3/16/13; 12128AA, exp. 4/6/13; 12150AA, exp. 4/28/13; 12191AA, exp. 6/8/13; 12212AA, exp. 6/29/13; 12226AA, exp. 7/13/13; 12248AA, exp. 8/4/13; 12261AA, exp. 8/17/13; 12282AA, exp. 9/7/13; 12303AA, exp. 9/28/13.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0626-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0625-2014Class IITerminatedVoluntary: Firm initiated
Z-0626-2014Class IITerminatedVoluntary: Firm initiated