Spacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor with...
Spacelabs Healthcare, Llc began a recall of Spacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor with arrhythmia detection or alarms on . FDA classified it as Class II and gives the reason as "Spacelabs Healthcare elance Vital Signs Monitor with Option S, model 93330, is recalled because the monitor will not permit connection of any of the Spacelabs TruLink line of SpO2 sensors". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0188-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Spacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor with arrhythmia detection or alarms.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Spacelabs Healthcare elance Vital Signs Monitor with Option S, model 93330, is recalled because the monitor will not permit connection of any of the Spacelabs TruLink line of SpO2 sensors.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Spacelabs Healthcare sent the "URGENT -MEDICAL DEVICE CORRECTION Spacelabs Healthcare elance Vital Signs Monitor with Option S" letter, dated 01 November 2013, to the US consignee on November 1, 2013. Spacelabs Healthcare will send the customer letter (translated as necessary) to all international subsidiaries and distributors of record via email on November 6, 2013. USERS ARE ADVISED THAT: -They may not be able to monitor their patient's oxygen saturation because the monitor will not allow the TruLink SpO2 sensor to be connected. The clinician will be required to find an alternate method to monitor oxygen saturation. -They will be contacted at the earliest possible date to schedule a convenient time and Spacelabs will correct all of your facility's affected monitors at no cost. Customer in the US can call Spacelabs Healthcare, Inc. of Snoqualmie, Washington at 800-522-7025. If you have any questions about this corrective action program, please contact Spacelabs Healthcare at +86 512 87178304 for Technical Support.
- Quantity in commerce
- 89 units total (1 in US and 88 International)
- Distribution
- Worldwide Distribution-Distributed in Puerto Rico and the countries of France, Indonesia, Korea, Nicaragua, Philippines, and Taiwan.
- Product codes and lots
- SERIAL NUMBER IN THE US: 3300-203410 SERIAL NUMBERS INTERNATIONAL: 3300-203361, 3300-203362, 3300-203363, 3300-203364, 3300-203386, 3300-203387, 3300-203388, 3300-203389, 3300-203390, 3300-203391, 3300-203392, 3300-203393, 3300-203394, 3300-203395, 3300-203396, 3300-203397, 3300-203398, 3300-203399, 3300-203400, 3300-203401, 3300-203402, 3300-203403, 3300-203404, 3300-203405, 3300-203406, 3300-203407, 3300-203408, 3300-203409, 3300-203410, 3300-203443, 3300-203444, 3300-203445, 3300-203446, 3300-203447, 3300-203448, 3300-203449, 3300-203450, 3300-203451, 3300-203452, 3300-203453, 3300-203454, 3300-203455, 3300-203456, 3300-203457, 3300-203458, 3300-203459, 3300-203460, 3300-203461, 3300-203462, 3300-203463, 3300-203464, 3300-203465, 3300-203466, 3300-203467, 3300-203468, 3300-203469, 3300-203470, 3300-203471, 3300-203472, 3300-203473, 3300-203474, 3300-203475, 3300-203476, 3300-203477, 3300-203478, 3300-203479, 3300-203480, 3300-203481, 3300-203482, 3300-203483, 3300-203484, 3300-203485, 3300-203486, 3300-203487, 3300-203488, 3300-203489, 3300-203490, 3300-203491, 3300-203492, 3300-203501, 3300-203502, 3300-203503, 3300-203504, 3300-203505, 3300-203506, 3300-203507, 3300-203508, 3300-203509, and 3300-203511.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0188-2014 | Class II | Terminated | Voluntary: Firm initiated |