Spacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmia...
Spacelabs Healthcare, Llc began a recall of Spacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmia detection or alarms. The qube is intended for use under... on . FDA classified it as Class II and gives the reason as "Spacelabs Healthcare qube Compact Patient Monitor, Model 91390, may fail to power ON due to incorrect programmed microcontroller component". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0351-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Spacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmia detection or alarms. The qube is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment. These devices determine a) when an alarm condition is violated; b) the alarm priority (i.e. high, medium or low); c) alarm limits; and d) when to initiate and terminate alarm notifications.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Spacelabs Healthcare qube Compact Patient Monitor, Model 91390, may fail to power ON due to incorrect programmed microcontroller component.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Spacelabs Healthcare sent the "URGENT -MEDICAL DEVICE CORRECTION Spacelabs Healthcare qube Compact Monitor" letter, dated 01 November 2013, to the US consignees/customers on November 1, 2013. Spacelabs Healthcare will send the customer letter (translated as necessary) to all international subsidiaries and distributors of record via email on November 6, 2013. The letter described the product, problem and actions to be taken. The customers were instructed to advise staff of the potential for the monitor to fail to turn ON. Spacelabs Healthcare will contact their customers at the earliest possible date to schedule a convenient time for Spacelabs to correct all of your facility's affected monitors at no cost. If you have any questions about this corrective action program, please contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
- Quantity in commerce
- 54 units total (42 units in the US and 12 outside US)
- Distribution
- Worldwide distribution: US (nationwide) including states of: Georgia, Idaho, New York, Oklahoma, Tennessee, Texas, and Wyoming and in countries of: Bahrain, Canada, France, Great Britain, Poland, and The Netherlands.
- Product codes and lots
- SERIAL NUMBERS IN THE US: 1390-101940, 1390-101945, 1390-101946, 1390-101947, 1390-101948, 1390-101950, 1390-101951, 1390-101952, 1390-101953, 1390-101954, 1390-101955, 1390-101956, 1390-101957, 1390-101958, 1390-101962, 1390-101963, 1390-101971, 1390-101972, 1390-101973, 1390-101974, 1390-101979, 1390-101980, 1390-101994, 1390-102002, 1390-102003, 1390-102004, 1390-102005, 1390-102006, 1390-102007, 1390-102008, 1390-102009, 1390-102010, 1390-102011, 1390-102012, 1390-102013, 1390-102014, 1390-102015, 1390-102016, 1390-102017, 1390-102018, 1390-102019, and 1390-102020. SERIAL NUMBER INTERNATIONAL: 1390-101949, 1390-101969, 1390-101970, 1390-101975, 1390-101976, 1390-101977, 1390-101978, 1390-102001, 1390-102021, 1390-102022, 1390-102023, and 1390-102024.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0351-2014 | Class II | Terminated | Voluntary: Firm initiated |