Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0838-2014Class II

DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Shoulder Blanket w/ Straps NS, REF T655NS, Rx only...

DeRoyal Industries, Inc. / initiated 2013-09-17 / Terminated / root cause: design

DeRoyal Industries Inc began a recall of DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Shoulder Blanket w/ Straps NS, REF T655NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane... on . FDA classified it as Class II and gives the reason as "The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0838-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
DeRoyal Industries, Inc.
Product
DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Shoulder Blanket w/ Straps NS, REF T655NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
Product code
ILO Pack, Hot Or Cold, Water Circulating
Reason for recall
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
Root cause tag
design
FDA root cause
Device Design
Action
DeRoyal sent a "URGENT VOLUNTARY RECALL DEROYAL HOT/COLD THERAPY UNIT" letter dated September 17, 2013. The letter identified the product, problem, and actions to be taken by the customers.
Quantity in commerce
2 units
Distribution
Nationwide Distribution.
Product codes and lots
Lot Number: 32871812

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0838-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0834-2014Class IITerminatedVoluntary: Firm initiated
Z-0835-2014Class IITerminatedVoluntary: Firm initiated
Z-0836-2014Class IITerminatedVoluntary: Firm initiated
Z-0837-2014Class IITerminatedVoluntary: Firm initiated
Z-0838-2014Class IITerminatedVoluntary: Firm initiated
Z-0839-2014Class IITerminatedVoluntary: Firm initiated