Z-0491-2014Class II

Carestream DRX-Revolution Mobile X-Ray System -- Made in U.S.A. by Carestream Health Inc. 150 Verona Street...

Eastman Kodak Company / initiated 2013-08-29 / Terminated

Carestream Health Inc. began a recall of Carestream DRX-Revolution Mobile X-Ray System -- Made in U.S.A. by Carestream Health Inc. 150 Verona Street, Rochester, NY . 14608 --- The DRX Revolution... on . FDA classified it as Class II and gives the reason as "It was discovered that Carestream received a report from a customer that 2 different patients were re-exposed due to the fact that the processed images were not able to be found in the system". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0491-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Eastman Kodak Company
Product
Carestream DRX-Revolution Mobile X-Ray System -- Made in U.S.A. by Carestream Health Inc. 150 Verona Street, Rochester, NY . 14608 --- The DRX Revolution Mobile system is a mobile diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. These products are mobile diagnostic x-ray systems designed to perform radiographic x-ray examinations on all pediatric and adult patients, in all patient treatment areas.
Product code
IZL System, X-Ray, Mobile
Reason for recall
It was discovered that Carestream received a report from a customer that 2 different patients were re-exposed due to the fact that the processed images were not able to be found in the system.
FDA root cause
Radiation Control for Health and Safety Act
Action
Carestream sent a letter to all affected customers informing them of the issue, how to detect the issue and actions to be taken by Carestream to correct the issue. The actions to be taken by Carestreamn to correct the issue will include the installation of a software patch to mitigate the effects of the observed failure and remove the possibility of a re-exposure in this failure mode. The correction will be implemented within the US to all affected devices at identified customer sites. For further questions call Carestream Customer Care at 1-800-328-2910, 8 am through 8 pm Eastern Standard Time.
Quantity in commerce
310 systerms domestically
Distribution
Nationwide
Product codes and lots
DRX Revolution Mobile X-Ray Systems manufactured from June 2012 through October 2013: Serial Numbers DRXR00690, 111, 114, 119, 121, 122, 123, 125, 126, 134, 142, 143, 151, 152, 153, 154, 155, 161, 162, 163, 164, 165, 166, 167, 168, 169, 173, 174, 176, 177, 178, 180, 181, 182, 183, 188, 189, 190, 191, 192, 196, 197, 198, 199, 201, 202, 203, 205, 206, 207, 208, 209, 211, 221, 222, 223, 224, 225, 226, 227, 228, 229, 230, 231, 232, 235, 236, 237, 238, 239, 240, 241, 242, 243, 244, 245, 246, 247, 248, 249, 254, 261, 262, 263, 264, 265, 266, 267, 268, 269, 270, 275, 276, 279, 280, 281, 282, 283, 284, 285, 286, 305, 312, 316, 317, 318, 319, 320, 321, 322, 323, 324, 325, 326, 330, 331, 332, 333, 334, 339, 340, 341, 342, 343, 344, 345, 346, 347, 348, 349, 353, 355, 357, 358, 359, 360, 361, 362, 363, 364, 365, 372, 373, 375, 376, 379, 380, 381, 382, 383, 384, 385, 386, 388, 394, 395, 396, 397, 398, 399, 400, 401, 403, 405, 412, 413, 414, 415, 417, 420, 421, 423, 431, 432, 433, 434, 435, 441, 442, 443, 444, 451, 452, 454, 455, 456, 457, 458, 459, 460, 461, 464, 466, 467, 487, 488, 489, 490, 495, 506, 510, 511, 512, 514, 515, 516, 517, 518, 519, 520, 521, 522, 523, 524, 530, 538, 541, 542, 543, 544, 545, 546, 547, 548, 549, 555, 557, 560, 561, 562, 563, 564, 581, 583, 584, 590, 591, 592, 593, 594, 595, 596, 597, 598, 599, 601, 602, 603, 609, 610, 611, 612, 613, 614, 616, 619, 620, 621, 622, 626, 627, 628, 629, 631, 632, 633, 634, 635, 636, 637, 639, 644, 645, 649, 650, 651, 656, 657, 658, 659, 660, 661, 663, 664, 665, 676, 677, 682, 685, 686, 687, 688, 689, 691, 692, 694, 696, 697, 698, 699, 700, 702, 703, 704, 705, 706, 707, 718, 723, 724.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0491-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0491-2014Class IITerminatedVoluntary: Firm initiated