Z-0389-2014Class II

Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms)

Datascope Corp. / initiated 2013-10-07 / Terminated / root cause: manufacturing process

Mindray DS USA, Inc. dba Mindray North America began a recall of Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) Product Usage: The V Series Monitor is intended for intra hospital use under the direct... on . FDA classified it as Class II and gives the reason as "Mindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily go blank, lose...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0389-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) Product Usage: The V Series Monitor is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The indications for use for the V Series include the monitoring of the following human physiological parameters: 1) ECG waveform derived from 3, 5, 6 and 12 lead measurements 2) Heart Rate 3) Pulse Oximtery 4) ST Segment 5) Arrhythmia Detection 6) Non Invasive Blood Pressure (IBP) 7) Cardiac Output (CO) 8) Respiratory Gases 9) Respiration Rate 10) Temperature The V series monitor has the capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices.
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Mindray has identified an issue with V Series Monitor where the V Dock power pins connecting the V Series Monitor to the V Dock may lose spring tension, potentially causing the screen to temporarily go blank, lose communication with the central station monitor (when one is in use) or shut down. No patient related events have been reported to Mindray concerning the issue.
Root cause tag
manufacturing process
FDA root cause
Environmental control
Action
Mindray sent an Urgent Corrective Action letter dated October 7, 2013 to all of their customers. The letter identified the affected product, problem and actions to be taken. Customers should contact their Mindray Service Representative at 1-800-288-2121 Monday through Friday from 830 am to 530 pm to arrange for the replacement of the affected product.
Quantity in commerce
1839 units
Distribution
Worldwide Distribution - USA Nationwide
Product codes and lots
V Series Montor p/n 0998-00-1801-01 and the associated V Dock , p/n 0998-UC-180-01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0389-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0389-2014Class IITerminatedVoluntary: Firm initiated