Z-0451-2014Class II

Abbott i-STAT PT/INR cartridges Abbott Point of Care Inc., Princeton, NJ 08540 The i-Stat PT a prothrombin...

Abbott Medical / initiated 2013-10-28 / Terminated / root cause: manufacturing process

Abbott Point Of Care Inc. began a recall of Abbott i-STAT PT/INR cartridges Abbott Point of Care Inc., Princeton, NJ 08540 The i-Stat PT a prothrombin time test is useful for monitoring patients... on . FDA classified it as Class II and gives the reason as "Abbott Point of Care Inc (APOC) has determined that i-Stat PT/INR cartridges have the potential to exhibit incorrectly elevated INR". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0451-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Abbott i-STAT PT/INR cartridges Abbott Point of Care Inc., Princeton, NJ 08540 The i-Stat PT a prothrombin time test is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or Warfarin. The cartridge is to be used with the i-Stat 1 Analyzer and as part of the i-Stat System, the PT/INR test is to be used by trained health care professionals in accordance with a facility's policies and procedures.
Product code
GJS Test, Time, Prothrombin
Reason for recall
Abbott Point of Care Inc (APOC) has determined that i-Stat PT/INR cartridges have the potential to exhibit incorrectly elevated INR.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Abbott Point of Care Inc. (APOC) issued an Urgent Recall Notice on October 28, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of cartridges from PT/INR lots in this letter. All unused cartridges would be credited and are to be returned as instructed on the enclosed Business Reply Card. If customers provided cartridges from the lot number listed above to another institution they were instructed to provide a copy of this letter to them. Customers with questions were instructerd to contact Abbott Point of Care Technical at 1-800-366-8020, Option 1. For questions regarding this recall call 800-366-8020.
Quantity in commerce
794,928 units US; 119,472 units OUS
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Brazil, Chile, China, Finland, France, Germany, Greece, Hong Kong, Ireland, Italy, South Korea, Kuwait, Mexico, Netherland, New Zealand, Nigeria, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, south Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE, and United Kingdom,
Product codes and lots
K020355  List number 04J50-01, 04J50-02; 03P89-24  Lot number R13149A; R13151; R13151A and cartridge lots from C13139 up to and including C13270A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0451-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0451-2014Class IITerminatedVoluntary: Firm initiated