SmartSite Low Sorbing Infusion Set with Texium Close Male Luer, Model No. 24301-0007T The SmartSite Low...
CareFusion 303, Inc. began a recall of SmartSite Low Sorbing Infusion Set with Texium Close Male Luer, Model No. 24301-0007T The SmartSite Low Sorbing Infusion Sets with Texium Male Luer are used to... on . FDA classified it as Class II and gives the reason as "CareFusion is recalling the SmartSite Low Sorbing Infusion Set with Texium Closed Male Luer because it is not assembled with low-sorbing tubing". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0450-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- SmartSite Low Sorbing Infusion Set with Texium Close Male Luer, Model No. 24301-0007T The SmartSite Low Sorbing Infusion Sets with Texium Male Luer are used to administer fluid and medications through a needle or catheter inserted into the patient's artery or vein.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- CareFusion is recalling the SmartSite Low Sorbing Infusion Set with Texium Closed Male Luer because it is not assembled with low-sorbing tubing.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- The firm, CareFusion, sent an "URGENT: Medical Device Recall Notification" letter dated November 8, 2013 to all customers who purchased the SmartSite Low Sorbing Infusion Set with Texium Closed Male Luer. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to immediately complete and return to CareFusion the enclosed, pre-addressed and postage paid, Recall Response Card and return all recalled product directly to the distributor form whom it was purchased. Customers with questions are instructed to contact their distributor directly or call the CareFusion Support Center at (888) 562-6018. If you have recall related questions or need support, please call CareFusion Support Center at 1-888-562-6018; hours: 7am to 4pm PST or Technical Support at 1-888-812-3229; hours 7am to 5pm PST.
- Quantity in commerce
- 480 units
- Distribution
- US Distribution in CA, WY, OH, UT, and MA.
- Product codes and lots
- Lot No: 13046215 13046641 13056285 13056577 13075931
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0450-2014 | Class II | Terminated | Voluntary: Firm initiated |