510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1215-2014Class II

Dental Abutment labeled in part: "4555 Riverside Drive Palm Beach Gardens, FL 33410 USA 1-800-342-5454***EDAZ...

ZimVie US Corp LLC / initiated 2013-08-26 / Terminated / root cause: design

Biomet 3i, LLC began a recall of Dental Abutment labeled in part: "4555 Riverside Drive Palm Beach Gardens, FL 33410 USA 1-800-342-5454***EDAZ , EDAZx and ILDExxx***BellaTek Zirconia Abutment... on . FDA classified it as Class II and gives the reason as "Reports relating to Biomet 3i's BellaTek Zirconia Dental Abutment fractures were received". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1215-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
Dental Abutment labeled in part: "4555 Riverside Drive Palm Beach Gardens, FL 33410 USA 1-800-342-5454***EDAZ , EDAZx and ILDExxx***BellaTek Zirconia Abutment Position Number: XX Abutment; Pilier; Abutment; Abutment; Pilastro; Pilar***Rx only***Non-sterile, Sterilize prior to use.***Biomet 3i Dental Iberica S.L.*** Biomet 3i Dental Abutments are intended for use as accessories to endosseous dental implants to support prostheses in edentulous or partially edentulous patients. These are intended for use to support single and multiple tooth prostheses.
Product code
NHA Abutment, Implant, Dental, Endosseous
Reason for recall
Reports relating to Biomet 3i's BellaTek Zirconia Dental Abutment fractures were received.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, BIOMET 3i, sent an "URGENT: Medical Device Removal - BellaTek Zirconia Abutments" letter dated August 2013 via email, facsimile or postal mail to its customers. The letter described the product, problem and actions to be taken. All customers were instructed to check their inventory for the affected product; return the abutments to BIOMET 3i for replacement and full credit; complete and return the attached Recall Return Response Form, and to advise their customers and forward a copy of the Medical Device Removal Notice and the contained recommendations. BIOMET 3i has decided to initiate a design change to mitigate the fracture risk, and consequently, will cease distribution of BellaTek Zirconia Abutments unit l the corrective actions are completed. If you have any further questions, please contact Customer Service at 1-800-342-5454 for assistance.
Quantity in commerce
10,454
Distribution
Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, Hawaii and Puerto Rico; and countries of: Spain, Australia, Canada, Colombia, Japan, Uruguay, Mexico, Estonia, United Kingdom, Denmark, Germany, Spain, Italy, Poland, Ireland Netherlands, France, Sweden, Portugal, Australia, France, Belgium, Switzerland, Norway, Czech Republic, France, Holland, and Luxembourg.
Product codes and lots
All lots of products made since April 1, 2011. Catalog model EDAZ, EDAZx and ILDExxx

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1215-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1215-2014Class IITerminatedVoluntary: Firm initiated