Z-0484-2014Class II

Helion S Exam Light or H300 The device is intended to locally illuminate an examination area of the patient's...

Hill-Rom Inc. / initiated 2013-02-19 / Terminated / root cause: manufacturing process

Trumpf Medical Systems, Inc. began a recall of Helion S Exam Light or H300 The device is intended to locally illuminate an examination area of the patient's body with high intensity light in a clinic or... on . FDA classified it as Class II and gives the reason as "Possible fatigue failure of the plastic joint may occur after an average use of seven years". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0484-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hill-Rom Inc.
Product
Helion S Exam Light or H300 The device is intended to locally illuminate an examination area of the patient's body with high intensity light in a clinic or doctor's practice.
Product code
FSY Light, Surgical, Ceiling Mounted
Reason for recall
Possible fatigue failure of the plastic joint may occur after an average use of seven years.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
TRUMPF sent an Urgent Recall Notice dated February 19, 2013, to all US affected customers. The letter identified the product the problem and the action needed to be taken by the customers. The recall notice instructed customers to verify if they still have these light systems, and if they do have the light system verify if they have a metal or plastic ring at the joint of concern (per the manufacturer's Safety Notice). TRUMPF will directly work with affected customers to replace the affected arm system. For further questions please call ( 888 ) 474-9359.
Quantity in commerce
287
Distribution
US Distribution including the states of AR, CA, CO, KS, MA, MI, MO, NJ, NY, NC, OH, OK, PA, SD, TX and WA.
Product codes and lots
Unknown

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0484-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0484-2014Class IITerminatedVoluntary: Firm initiated