REF 200068 Security (TM) Enclosed Carpal Tunnel System Blade ( Single) ABS /Stainless Steel Sterile Device...
Biomet, Inc. began a recall of REF 200068 Security (TM) Enclosed Carpal Tunnel System Blade ( Single) ABS /Stainless Steel Sterile Device, Single Use Only. Designed to release the transverse... on . FDA classified it as Class II and gives the reason as "Security Blades were manufactured oversized and potentially won't advance or may be difficult to advance through the mating Security Clip". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0381-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- REF 200068 Security (TM) Enclosed Carpal Tunnel System Blade ( Single) ABS /Stainless Steel Sterile Device, Single Use Only. Designed to release the transverse carpal ligament.
- Product code
- FZX Guide, Surgical, Instrument
- Reason for recall
- Security Blades were manufactured oversized and potentially won't advance or may be difficult to advance through the mating Security Clip.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- 10/25/2013 Urgent Medical Device Recall Notifications were sent to each Distributor and Risk/Recall Manager at each direct account identifying the recalled device, possible adverse health consequences and instructing consignees to immediately locate and remove devices from circulation, following instructions on the FAX back form and returning it within 3 business days to FAX 574-372-1683 and returning product to Biomet Return Goods, 56 East Bell Drive Warsaw In 46580. Questions should be directed to 574-372-1570 M-F, 8 a.m. to 5 p.m.
- Quantity in commerce
- 1130
- Distribution
- Worldwide Distribution: USA (nationwide) and Internationally to Japan.
- Product codes and lots
- Catalog Number: 200068 Medical Device Listing Number: B008657 Lot Number Identification: 088610, 090600, 143810, 257060, 289900, 365910, 460490, 558300, 626660, 697320,767260,795090,857730,864640,979680
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0381-2014 | Class II | Terminated | Voluntary: Firm initiated |