Z-0456-2014Class II

iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological

IBA Dosimetry GmbH / initiated 2013-08-20 / Terminated / root cause: software

Iba Dosimetry Gmbh began a recall of iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological on . FDA classified it as Class II and gives the reason as "The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0456-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
IBA Dosimetry GmbH
Product
iba COMPASS SW Version 3.0, Catalog Number CS10-100 (medical linear accelerator) radiological
Product code
IYE Accelerator, Linear, Medical
Reason for recall
The development partner of IBA informed that there is problem in a SW component used in COMPASS when converting Region of Interest (ROI) geometries to dose grid representation. Under certain conditions this problem generates too large ROI voxel volumes causing erroneous dose statistics and DVHs.
Root cause tag
software
FDA root cause
Software design
Action
Iba sent an Field Safety Notice on August 20, 2013, by email and letter delivered by UPS, Please distribute this notice to all those who need to be aware within your area of responsibility where the affected software have been installed (e.g. colleagues, satellites). Please maintain awareness of this notice and resulting action until we have provided you with a new software version. We would kindly ask you to acknowledge the receipt of this Field Safety Notice. Please sign and send back the attached Confirmation of Receipt within four weeks after the receipt of this Field Safety Notice. We deeply apologize for the inconvenience. We are working on a new software version with corrected functionality. This should become available during next fall and we will ship it to you as soon as it is released, together with the installation and verification instructions. Should you have questions or require additional information, please contact the Customer Service Team at: Service Department IBA Dosimetry GmbH Bahnhofstrasse 5 DE-90592 Schwarzenbruck, Germany Tel.:+49 9128 607 38 Fax: + 49 9128 607 26 Email: [email protected] .
Quantity in commerce
33 units
Distribution
Worldwide Distribution - US Distribution including the states of CT, MA, OH, CA, SC, NY and OK, and the countries of International distribution (from Germany) Poland, Germany, GB, The Netherlands, Spain, Turkey, Canada, Mexico, China, Hong Kong and Singapore.
Product codes and lots
SW Version 3.0. Catalong Number CS10-100

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0456-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0456-2014Class IITerminatedVoluntary: Firm initiated