Z-0905-2014Class II

Medtronic Advanced Energy Aquamantys3 Pump Generator; Product Catalog Number: 10-1357 (Generator assembly)...

Medtronic Navigation, Inc / initiated 2013-11-12 / Terminated / root cause: design

Medtronic Advanced Energy, LLC began a recall of Medtronic Advanced Energy Aquamantys3 Pump Generator; Product Catalog Number: 10-1357 (Generator assembly), 40-404-1(as shipped, including accessories) The... on . FDA classified it as Class II and gives the reason as "The internal protection circuitry on the electrosurgical output that prevents a patient from exposure to DC voltage when the device is activated is being compromised during normal use". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0905-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Navigation, Inc
Product
Medtronic Advanced Energy Aquamantys3 Pump Generator; Product Catalog Number: 10-1357 (Generator assembly), 40-404-1(as shipped, including accessories) The Aquamantys3 System combines radio frequency (RF) energy and saline to reduce blood loss during and after surgical procedures. This patented Transcollation technology has been shown to reduce transfusion rates and may also reduce the need for other blood management procedures.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
The internal protection circuitry on the electrosurgical output that prevents a patient from exposure to DC voltage when the device is activated is being compromised during normal use.
Root cause tag
design
FDA root cause
Device Design
Action
Medtronic sent an Urgent Product Removal Notification on November 12, 2013, via Next Day FEDEX. The communication advises users to immediately stop using the affected AQM3 electrosurgical generators and quarantine them until they can be returned to Medtronic Advanced Energy. Medtronic field personnel will collect them in order to remove them from service and return them to Medtronic Advanced Energy. Customers with questions were instructed to contact Customer Service at 866-777-9400. For questions regarding this recall call 866-777-9400.
Quantity in commerce
242 devices
Distribution
Nationwide Distribution including CA, MO, PA, NJ, NY, OH, VA, TX, NC, DE, KS, TN, LA, IL, MI, SC, WI, and FL.
Product codes and lots
lowest serial number is GN001141; the highest is GN001704 (non-consecutive)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0905-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0905-2014Class IITerminatedVoluntary: Firm initiated