FastPack Total PSA Immunoassay Chemilunescence assay for the determination of Total PSA. FastPack Total PSA...
Qualigen Inc began a recall of FastPack Total PSA Immunoassay Chemilunescence assay for the determination of Total PSA. FastPack Total PSA Immunoassay is a paramagnetic particle immunoassay... on . FDA classified it as Class II and gives the reason as "Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0547-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Qualigen Inc
- Product
- FastPack Total PSA Immunoassay Chemilunescence assay for the determination of Total PSA. FastPack Total PSA Immunoassay is a paramagnetic particle immunoassay for the in vitro quantitative determination of prostate-specific antigen (PSA) in human serum and plasma as an aid in the management of patients with prostate cancer. The FastPack Total PSA Immunoassay is designed for use with the FastPack System.
- Product code
- LTJ Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers
- Reason for recall
- Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- Qualigen, Inc., notified customers of the recall by calling customers via telephone (with a telephone script) on 11/08/2013. The telephone (script) included the reason for recall, product description with codes, instructions and contact information. The customers were instructed to examine your inventory and discontinue use of the kits immediately. Qualigen will send you replacement kits and provide you with a shipping label to return all of the defective kits including any that were partially used. Feel free to contact System Support at 760.579.6900 if you have any questions or require assistance with returning your kits.
- Quantity in commerce
- 182
- Distribution
- US Distribution to states of: AL, CA, KS, KY, IN, MA, NC, NY, NV, OK, SC, TN, TX, UT, VA, and WA.
- Product codes and lots
- Total PSA: Catalog Number 25000040. Lot Number 1309038-10P Exp Date 6/26/2013.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0547-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0548-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0549-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0550-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY