Z-0548-2014Class II

FastPack Testosterone Immunoassay Chemilunescence assay for the determination of Testosterone The FastPack...

Qualigen Inc / initiated 2013-11-08 / Terminated / root cause: manufacturing process

Qualigen Inc began a recall of FastPack Testosterone Immunoassay Chemilunescence assay for the determination of Testosterone The FastPack Testo Immunoassay is a paramagnetic particle... on . FDA classified it as Class II and gives the reason as "Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0548-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Qualigen Inc
Product
FastPack Testosterone Immunoassay Chemilunescence assay for the determination of Testosterone The FastPack Testo Immunoassay is a paramagnetic particle immunoassay for the in vitro quantitative determination of total testosterone in human serum and plasma. It is intended strictly for in-vitro diagnostic use as an aid in the diagnosis and management of conditions involving excess or deficiency of this androgen. The FastPack Testo Immunoassay is designed for use with the FastPack System.
Product code
CDZ Radioimmunoassay, Testosterones And Dihydrotestosterone
Reason for recall
Qualigen initiated this recall because the FastPack kit may cause lower than expected results due to a manufacturing defect.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
Qualigen, Inc., notified customers of the recall by calling customers via telephone (with a telephone script) on 11/08/2013. The telephone (script) included the reason for recall, product description with codes, instructions and contact information. The customers were instructed to examine your inventory and discontinue use of the kits immediately. Qualigen will send you replacement kits and provide you with a shipping label to return all of the defective kits including any that were partially used. Feel free to contact System Support at 760.579.6900 if you have any questions or require assistance with returning your kits.
Quantity in commerce
18
Distribution
US Distribution to states of: AL, CA, KS, KY, IN, MA, NC, NY, NV, OK, SC, TN, TX, UT, VA, and WA.
Product codes and lots
Testosterone. Catalog Number 25000041. Lot Number 1310019-3P Exp Date: 9/22/2014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0548-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0547-2014Class IITerminatedVoluntary: Firm initiated
Z-0548-2014Class IITerminatedVoluntary: Firm initiated
Z-0549-2014Class IITerminatedVoluntary: Firm initiated
Z-0550-2014Class IITerminatedVoluntary: Firm initiated