Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device...
Alere San Diego, Inc. began a recall of Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device with BZO, Reditest... on . FDA classified it as Class II and gives the reason as "Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0653-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Alere San Diego, Inc.
- Product
- Product Brand Name(s): Innovacon Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device, OrALert Oral Fluid Drug Screen Device with BZO, Reditest On-Site Oral Fluid Drug Screen Device & XALEX One Step Rapid Drug Test. Type of Packaging: 25 individually pouched test devices & collectors in a labeled kit box. The OrALert Oral Fluid Drug Screen Device for AMP/mAMP/COC/OPI/THC/PCP/BZO is an immunoassay based on the principle of competitive binding. Drugs that may be present in the oral fluid specimen compete against their respective drug conjugates for binding sites on their specific antibody.
- Product code
- DKZ Enzyme Immunoassay, Amphetamine
- Reason for recall
- Alere initiated this recall of certain lots of the Oral Fluid Drug Screen Device because the sponge may become dislodged from the Saliva Collector handle prior to or during the oral fluid sample collection process. If the sponge becomes dislodged during the specimen collection process, the donor may inadvertently swallow the sponge which could present a choking risk.
- FDA root cause
- Other
- Action
- Firm sent customers notification letter via combination of fax, email or direct mail. The notification letter dated June 3, 2013, titled "URGENT PRODUCT SAFETY NOTICE", informed customers of the recall, reason for recall, product description with lot codes, potential hazard statement, customer required actions, and contact information.
- Quantity in commerce
- 280,710
- Distribution
- US Distribution: including states of: AL, CA, PA, and VA.
- Product codes and lots
- Lot/Unit Codes: DSD-765-011 (Alere PN 4012350099) DOA2100709, DOA2110198, DOA2120093, DOA3010031, DOA3010095, DOA3010096, DOA3010393, DOA3010566, DOA3020420. DSD-765-011 (Alere PN 4080350099) DOA2100672, DOA2120094, DOA2120095, DOA2120158, DOA3010013, DOA3010064, DOA3010097, DOA3010394, DOA3010565, DOA3020557. DSF-765-031 (Alere PN 4584150099) DOA2100438, DOA2100667, DOA2110073, DOA2110315, DOA2120096, DOA2120738, DOA2120739, DOA3010003, DOA3010004, DOA3010005, DOA3010199, DOA3010202, DOA3020182, DOA3020183. DSF-765-011 (Alere PN 4582290164) DOA2100250, DOA2110316 , DOA2110317 , DOA2120502, DOA2120503, DOA3010236. DSF-765-011 (Alere PN 4028220211) DOA2100614. Expiration date, or Expected shelf life: 10/31/14 to 2/28/15. Expected shelf life: 24 months.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0653-2014 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY