AMSCO Evolution steam sterilizer Transfer Carriage
Steris Corporation began a recall of AMSCO Evolution steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into... on . FDA classified it as Class II and gives the reason as "The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0478-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- AMSCO Evolution steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into and out of steam sterilizer units.
- Product code
- FLE Sterilizer, Steam
- Reason for recall
- The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Steris sent an Urgent Voluntary Field Correction Notice letter dated October 18, 2013 to their customers. The letter identified the affected product, problem, and actions to be taken. The letter informed customers that a Steris Service Representative will contact affected facility to arrange for Steris to upgrade the Transfer Carriage(s). For questions contact Steris Field Service Dispatch at 1-800-333-8828.
- Quantity in commerce
- 936 units
- Distribution
- Worldwide Distribution - USA Nationwide in the states of: AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MI, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and the District of Columbia including the countries of: Brazil, Canada, Chile, Guam, Mexico, Saudi Arabia, United Arab Emirates & Venezuela.
- Product codes and lots
- Serial Number Range: 030061103 - 0535531220
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0478-2014 | Class II | Terminated | Voluntary: Firm initiated |