Z-0478-2014Class II

AMSCO Evolution steam sterilizer Transfer Carriage

STERIS Corporation / initiated 2013-10-18 / Terminated / root cause: design

Steris Corporation began a recall of AMSCO Evolution steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into... on . FDA classified it as Class II and gives the reason as "The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0478-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
AMSCO Evolution steam sterilizer Transfer Carriage. Product Usage: Mobile, wheeled transportation unit utilized to carry, load, & unload the Loading Car into and out of steam sterilizer units.
Product code
FLE Sterilizer, Steam
Reason for recall
The firm discovered that unlatching can occur when overloaded transfer carriages are pulled or maneuvered incorrectly or with excessive force.
Root cause tag
design
FDA root cause
Device Design
Action
Steris sent an Urgent Voluntary Field Correction Notice letter dated October 18, 2013 to their customers. The letter identified the affected product, problem, and actions to be taken. The letter informed customers that a Steris Service Representative will contact affected facility to arrange for Steris to upgrade the Transfer Carriage(s). For questions contact Steris Field Service Dispatch at 1-800-333-8828.
Quantity in commerce
936 units
Distribution
Worldwide Distribution - USA Nationwide in the states of: AL, AZ, AR, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, ME, MD, MA, MI, MN, MI, MO, MT, NE, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY and the District of Columbia including the countries of: Brazil, Canada, Chile, Guam, Mexico, Saudi Arabia, United Arab Emirates & Venezuela.
Product codes and lots
Serial Number Range: 030061103 - 0535531220

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0478-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0478-2014Class IITerminatedVoluntary: Firm initiated