Z-0533-2014Class II

Integra Disposable, Convenience Kit Single Shot Epidural Tray Reorder Number 340269 Rx Only

Integra Pain Management / initiated 2013-10-28 / Terminated / root cause: design

Integra LifeSciences Corp. d.b.a. Integra Pain Management began a recall of Integra Disposable, Convenience Kit Single Shot Epidural Tray Reorder Number 340269 Rx Only Product Usage: The Integra kits which contain a BD LOR syringe are... on . FDA classified it as Class II and gives the reason as "Integra LifeSciences Corporation is recalling certain lots of Integra Kits for Epidural Anesthesia Administration as a result of a Product Advisory Notice from BD Medical for the LOR syringes indicating the plunger of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0533-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra Pain Management
Product
Integra Disposable, Convenience Kit Single Shot Epidural Tray Reorder Number 340269 Rx Only Product Usage: The Integra kits which contain a BD LOR syringe are intended to provide licensed medical professionals with a kit containing drugs and other medical devices that are used to support the administration of general, local or spinal anesthetics. The BD LOR syringe is intended for use, in conjunction with an epidural needle, to verify the needle tip placement is in the epidural space by use of the Loss of Resistance Technique as detailed in medical textbooks and medical journal articles. The LOR Syringe is not intended for injection or aspiration.
Product code
CAZ Anesthesia Conduction Kit
Reason for recall
Integra LifeSciences Corporation is recalling certain lots of Integra Kits for Epidural Anesthesia Administration as a result of a Product Advisory Notice from BD Medical for the LOR syringes indicating the plunger of the syringe may stall or stick when travelling within the syringe.
Root cause tag
design
FDA root cause
Component design/selection
Action
BD Medical sent a Product Advisory Notice letter dated August 23, 2013 to Iaffected consignees by via traceable courier. The letter identified the affected product, problem and actions to be taken. Distributor have been asked to forward copies of the notification to their customers. For questions call BD Customer Service at 1-888-237-2762.
Quantity in commerce
359,250
Distribution
USA Nationwide Distribution and one foreign consignee in Mursalat.
Product codes and lots
Kit Cat No: 10-2433,13-2595,13-2616,13-2628,3104195,3403061,3403150,3403316,3403579,3403727,3403776,3403813,3403989,3404131,3404269,3404293,3404417,3404468,3404578,3503559,CUS025,CUS1121-01,CUS1164-02,CUS1202-04,CUS1298-03,CUS1376,CUS1376,CUS1400-01,CUS1406-03,CUS1415-04,CUS1568-01,CUS1664-01,CUS1696,CUS1723,CUS1760,CUS1773-01,CUS1797,CUS279-02,CUS416,CUS597,CUS628-02,CUS755,PISLL7,RSM394,RSM395,SS1246,SS2036-01,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0533-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0533-2014Class IITerminatedVoluntary: Firm initiated