Z-1068-2014Class II
Renasys EZ PLUS, a Canister Component of of Renasys EZ Negative Pressure Wound Therapy System
Smith & Nephew, Inc. / initiated 2013-11-14 / Terminated / root cause: design
Smith & Nephew Inc. began a recall of Renasys EZ PLUS, a Canister Component of of Renasys EZ Negative Pressure Wound Therapy System on . FDA classified it as Class II and gives the reason as "Modification of the bacterial overflow guard (filter) and related changes to product labeling". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1068-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- Renasys EZ PLUS, a Canister Component of of Renasys EZ Negative Pressure Wound Therapy System
- Product code
- OMP Negative Pressure Wound Therapy Powered Suction Pump
- Reason for recall
- Modification of the bacterial overflow guard (filter) and related changes to product labeling.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The correction strategy is as follows: " An Urgent Medical Device Correction Notice has been prepared to be distributed to all US consignees via Federal Express delivery for confirmation of the delivery. " A copy of the Urgent Medical Device Correction Notice is to be posted on Smith & Nephew's websites, URL corporate, http://www.smith-nephew.com/news-and-media/, and myrenasys.com, concurrent with the distribution of the Notice to consignees. " US distribution of the original design of Canister Kits codes (66800912, 66800913, 66801066, 66800423 and 66800058) subject to this correction ceased November 1, 2013. " The correction has been implemented on a change being affected basis. The modified design is being distributed to US customers effective November 1, 2013. No product will be removed from the market. " Periodic progress reports to FDA, as required per 21 CFR 806 will be submitted monthly until completion of the correction actions..
- Quantity in commerce
- 6,871,375 canisters
- Distribution
- Worldwide Distribution: USA (nationwide) including states of: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NH, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, VA, WA, WV, and Puerto Rico; and the countries of: Austria, Australia, Belgium, Switzerland, China, Cyprus, Germany, Denmark, Spain, Finland, France, UK, Greece, Israel, Italy, Malta, Netherlands, Norway, Paraguay, Russia, Saudi Arabia, Sweden, Thailand, Turkey, South Africa, Mexico, and Canada.
- Product codes and lots
- Product code: 66800912 - 800ml canister with solidifier 66800913 - 250ml canister with solidifier 66801066 - 800ml canister without solidifier 66800423 - 800ml canister with solidifier 66800058 - 250ml canister with solidifier Lot codes - M201498 thru M400299
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1067-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1068-2014 | Class II | Terminated | Voluntary: Firm initiated |