Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0516-2014Class II

HemosIL High Abnormal Control 3 UNASSAYED Part Number: 0020003320

Instrumentation Laboratory Co. / initiated 2013-11-15 / Terminated

Instrumentation Laboratory Co. began a recall of HemosIL High Abnormal Control 3 UNASSAYED Part Number: 0020003320 on . FDA classified it as Class II and gives the reason as "HemosIL Low, Abnormal, Normal Controls UNASSAYED (no-value assigned) lacks 510(k) approval". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0516-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Instrumentation Laboratory Co.
Product
HemosIL High Abnormal Control 3 UNASSAYED Part Number: 0020003320
Product code
GGN Plasma, Coagulation Control
Reason for recall
HemosIL Low, Abnormal, Normal Controls UNASSAYED (no-value assigned) lacks 510(k) approval
FDA root cause
No Marketing Application
Action
Instrumentation Laboratory issued Important Correction Notice via USPS to US customers dated November 2013 advising users in the US that unassayed controls are not acceptable for use in their unassayed (non-value assigned) format. IL will no longer sell or distribute controls without assayed values for the US market as of November 15, 2013, and are converting US customers to an assayed control format. Additional contact with customers via phone calls and a reminder email. An enclosed response tracking form provided to be returned. For technical questions, contact the IL Technical Support Center at 1-800-678-0710, Option # 6 (ACL Coagulation Products). For replacement product if value assignments are not available for your product lot, please contact your IL Customer Service Representative at 1-800-955-9525, Option # 6. For questions regarding this recall call 781-861-4467.
Quantity in commerce
84561 kits
Distribution
Nationwide Distribution
Product codes and lots
All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0516-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0514-2014Class IITerminatedVoluntary: Firm initiated
Z-0515-2014Class IITerminatedVoluntary: Firm initiated
Z-0516-2014Class IITerminatedVoluntary: Firm initiated