Asante Conset Infusion Set with 60cm tubing & 9mm cannula packaged in 5-packs,singles, or inside a...
Asante Solutions, Inc. began a recall of Asante Conset Infusion Set with 60cm tubing & 9mm cannula packaged in 5-packs,singles, or inside a Combination Kit (FG-5229-4); Distributed by Asante... on . FDA classified it as Class II and gives the reason as "Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0599-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Asante Solutions, Inc.
- Product
- Asante Conset Infusion Set with 60cm tubing & 9mm cannula packaged in 5-packs,singles, or inside a Combination Kit (FG-5229-4); Distributed by Asante Solutions, Sunnyvale, CA; Manufactured by Convatec; Lejre, Denmark.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Asante sent urgent medical device recall notifications letters to distributors and customers on November 19, 2013 by US mail, return receipt requested.
- Quantity in commerce
- 1979 kits
- Distribution
- Nationwide Distribution.
- Product codes and lots
- Asante Catalog number:FG-4229-5 FG-4229T; Unomedical Item number:43-060-5269; Lot numbers: 0230127 5005802 5038313.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0594-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0595-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0596-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0597-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0598-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0599-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0600-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0601-2014 | Class II | Terminated | Voluntary: Firm initiated |