Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0965-2014Class III

D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories...

Bio-Rad Laboratories Inc. / initiated 2013-11-22 / Terminated / root cause: software

Bio-Rad Laboratories, Inc. began a recall of D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin... on . FDA classified it as Class III and gives the reason as "D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0965-2014
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography.
Product code
LCP Assay, Glycosylated Hemoglobin
Reason for recall
D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System.
Root cause tag
software
FDA root cause
Software design
Action
BIO-RAD sent an Urgent Medical Device Correction letter dated November 22, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to continue to use the affected update kit disk(s) included with the reorder pack(s), but manually track the number of injections on the cartridge, reviewing QC results closely throughout the life of the reorder pack or destroy the affected disk(s) included with these reorder pack(s) and replace with the new Update Kit disk(s) The new Update Kit disks will elminate the problem and may be used with all reorder packs. The replacement disks do not contain reagent or calibrator information; after installing teh new Update Kit disk, they ust manually update the component information using instructions provided with the disk. For questions regarding this recall call 510-741-6265.
Quantity in commerce
2906 kits (318 Kits in US; 2588 kits internationally)
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Latin America, Asia Pacific, Canada, Australia, New Zealand and India.
Product codes and lots
Catalog number: 220-0201: Reorder Pack Lot No./Expiration date: 20030322 3/31/2014 20030337 3/31/2014 64000081 5/31/2014  64000331 5/31/2014  64000332 5/31/2014 64000888 08/31/2014 64001344 04/30/2014 64002239 11/30/2014   ( 64002240 08/3112014  64002241 11/30/2014

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0965-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0965-2014Class IIITerminatedVoluntary: Firm initiated