MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic...

Roche Diagnostics GmbH / Terminated / root cause: software

Roche Diagnostics Operations, Inc. began a recall of MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from different kinds of crude sample.... FDA classified it as Class II and gives the reason as "This issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a potential for sample mismatch when using the MagNA Pure 96 IVD...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0536-2014
Classification
Class II
Status
Terminated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from different kinds of crude sample material (whole blood, serum, blood cells, culture cells, tissue, bacteria, fungi) by the use of the specially designed MagNA Pure reagent kits, automated filling of different kinds of PCR reaction vessels.
Product code
JJH Clinical Sample Concentrator
Reason for recall
This issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a potential for sample mismatch when using the MagNA Pure 96 IVD instrument (catalog number 06541089001). Importing an order file in xml format to the MagNA Pure 96 IVD instrument software may result in a wrong sequence of samples in the graphical overvi
Root cause tag
software
FDA root cause
Software design
Action
Roche Diagnostics sent an URGENT Medical Device Correction Notice dated November 25, 2013, to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers were advised to discontinue use of the Pipetting Report and to use the sample list file instead. The MagNA Pure 96 Instrument Operators Guide is being updated with an addendum addressing this issue. The letter included a Customer Response Form for customers to complete and return to the firm by FAX : 1-888-670-4135. Please contact the Roche Support Network Customer Support Center at 1-800-526-1247, 24 hours a day, seven days a week, if you have questions about the information contained in this UMDC.
Quantity in commerce
27
Distribution
US Distribution including the states of AL, AZ, CA, CT, GA, NY, NC, NJ, OH, TX, VA and WA.
Product codes and lots
Serial Numbers 2001, 2002, 2003, 2004, 2006, 2011, 2012, 2013, 2014, 2015, 2016, 2018, 2019, 2020, 2021, 2022, 2023, 2024, 2025, 2026, 2027, 2028, 4001, 4022, 4023, 4024, 4025.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0536-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0536-2014Class IITerminatedVoluntary: Firm initiated