510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0855-2014Class II

Orthofix Firebird Spinal Fixation System Instrumentation Modular Screw Driver

Orthofix SRL / initiated 2013-11-26 / Terminated / root cause: manufacturing process

Orthofix, Inc. began a recall of Orthofix Firebird Spinal Fixation System Instrumentation Modular Screw Driver, Product Usage: The Modular Screw Driver is used for inserting a modular bone... on . FDA classified it as Class II and gives the reason as "Orthofix received 6 complaints which resulted in a reportable events due to extended surgical times greater than 30 minutes for the Modular Screw Driver (PN 52-1332). The complaints alleged that the screw driver's...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0855-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthofix SRL
Product
Orthofix Firebird Spinal Fixation System Instrumentation Modular Screw Driver, Product Usage: The Modular Screw Driver is used for inserting a modular bone screw into the pedicle during a spinal fixation procedure. It is a reusable instrument.
Product code
KWP Appliance, Fixation, Spinal Interlaminal
Reason for recall
Orthofix received 6 complaints which resulted in a reportable events due to extended surgical times greater than 30 minutes for the Modular Screw Driver (PN 52-1332). The complaints alleged that the screw driver's collet would malfunction resulting in the surgeon being unable to use the Modular Screw Driver to effectively place Modular Screws, which may result in a delay of surgery.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Orthofix sent an Urgent Medical Device Recall Notification letter dated November 26, 2013 by certified mail to distributors. The letter identified the affected product, problem and actions to be taken. The letter instructed distributors to immediately cease any further distribution, return affected products to Orthofix at no cost and complete the attached reply form. For questions contact your local Orthofix representative.
Quantity in commerce
443
Distribution
Worldwide Distribution - US Nationwide including (Puerto Rico) and the countries of: Spain, Germany, Australia, Peru, Mexico, Colombia, South Africa, and Israel.
Product codes and lots
Production Identification Numbers: Part Number 52-1332, All Lot Numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0855-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0855-2014Class IITerminatedVoluntary: Firm initiated