ABACUS TPN Calculation Software, Baxa: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE...

Baxter Corporation Englewood / initiated 2013-12-03 / Terminated / root cause: software

Baxter Corporation Englewood began a recall of ABACUS TPN Calculation Software, Baxa: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME... on . FDA classified it as Class II and gives the reason as "Baxter Healthcare Corporation is issuing a voluntary correction for ABACUS Total parenteral Nutrition (TPN) Calculation Software versions 2.0 through 3.0 due to the possibility that a patient could receive another...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0539-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Corporation Englewood
Product
ABACUS TPN Calculation Software, Baxa: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME (Multi-Workstation Edition) ABACUS is a Windows - based order entry TPN software application for comprehensive calculations and label printing. The primary use of Abacus is for TPN order calculations. Other uses of Abacus include: non-TPN volumes administered to patients on TPN therapy in order calculations; CVVH calculations; Cardioplegia (Buckburg) calculations; Epidural calculations; Dilutions from concentrated ingredient; Combinations of multiple ingredients; Fractional doses from larger volumes
Product code
LNX
Reason for recall
Baxter Healthcare Corporation is issuing a voluntary correction for ABACUS Total parenteral Nutrition (TPN) Calculation Software versions 2.0 through 3.0 due to the possibility that a patient could receive another patient's TPN order as a result of data corruption.
Root cause tag
software
FDA root cause
Unknown/Undetermined by firm
Action
Baxter sent an Urgent Device Correction letter dated December 4, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check the patient order against the final bag label as part of the TPN acceptance procedure to ensure that the patient's TPN order is correct. Customers will be contacted by Baxter to arrange for installation of a software upgrade specific to their compounder's configuation. Customers were asked to acknowledge the receipt of this notification by completing the attached Customer Reply Form and return it to Baxter by fax to 1-224-270-5457 or scanning and emailing it to [email protected]. If the product was distributed to other facilities customers were instructed to notify end-user customers in accordance with their customary procedures. Customers with questions were instructed to call 1-800-678-2292. For questions regarding this recall call 800-422-9837.
Quantity in commerce
2,211
Distribution
Nationwide Distribution
Product codes and lots
Product Code: 8300-0045,,8300-0046,,8300-0047,,8300-0045,,8300-0046,,8300-0047,,8300-0045,,8300-0046,,8300-0047,,8300-0045,,8300-0046,,8300-0047,,8300-0045,,8300-0046,,8300-0047

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0539-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0539-2014Class IITerminatedVoluntary: Firm initiated