Sandel Healthcare Safety Devices Weighted Safety Scalpel #3 Disposable Safety Handle with #15 Blade...
Ansell Sandel Medical Industries LLC began a recall of Sandel Healthcare Safety Devices Weighted Safety Scalpel #3 Disposable Safety Handle with #15 Blade Manufactured for: Ansell Sandel Medical Solutions LLC 19736... on . FDA classified it as Class II and gives the reason as "Select lots of SANDEL Weighted Safety Scalpels are packaged with the scalpel blade exposed beyond the safety shield which could compromise the integrity of the package and the sterility of the scalpel leading to injury". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0639-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ansell Sandel Medical Solutions LLC
- Product
- Sandel Healthcare Safety Devices Weighted Safety Scalpel #3 Disposable Safety Handle with #15 Blade Manufactured for: Ansell Sandel Medical Solutions LLC 19736 Dearborn St. Chatsworth, CA 91311 USA Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands Made in China Packaged in Mexico Ansell A manual surgical scalpel blade for making incisions in various general surgical procedures.
- Product code
- GDZ Handle, Scalpel
- Reason for recall
- Select lots of SANDEL Weighted Safety Scalpels are packaged with the scalpel blade exposed beyond the safety shield which could compromise the integrity of the package and the sterility of the scalpel leading to injury.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- The firm, Ansell, sent an "URGENT: VOLUNTARY RECALL OF ANSELL SANDEL WEIGHTED SAFET SCALPEL" letter dated November 11, 2013 to its customers and posted a website notification on November 11, 2013. The letter described the product, problem and actions to be taken. The customers were instructed: Do not ship SANDEL Weighted Safety Scalpel products 2210-L, 2210-N, 2211-L, 2211-N, 2215-L, 2215-N, or 2220-N to a hospital or end user with the listed lot numbers; quarantine SANDEL Weighted Safety Scalpel products in your inventory; and complete the attached response form and return it to attention: Ansell; via fax at 818-534-2510 or email [email protected]. Additionally, if you have distributed this product to other accounts, please notify them of the voluntary recall and instruct them to return any outstanding stock to you as indicated in the attachment. If you have any questions or need additional information, please contact Customer Service by calling 866-764-3327 (6:00am-5:00pm PST/9:00am-8pm EST) or email [email protected]
- Quantity in commerce
- 1203 - 96 cases (4128 units) ; 1553 -96 cases (3548 units)
- Distribution
- Worldwide Distribution: US (nationwide) including states of: AL, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MN, MS, NC, NE, NY, OK, PA, PR, RI, TN, TX and WI; and country of: Canada.
- Product codes and lots
- 510 K exempt Device Listing Number D129456 Model Number 2215-N LN 1203 and 1553
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0635-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0636-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0637-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0638-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0639-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0640-2014 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0641-2014 | Class II | Terminated | Voluntary: Firm initiated |