Z-0584-2014Class II

Philips Essenta DR, a Stationary x-ray system, X-ray system Model: 712070 The Essenta DR is a multifunctional...

Philips North America LLC / initiated 2013-12-05 / Terminated / root cause: manufacturing process

Philips Medical Systems, Inc. began a recall of Philips Essenta DR, a Stationary x-ray system, X-ray system Model: 712070 The Essenta DR is a multifunctional stationary X-ray system, in which the swivel arm... on . FDA classified it as Class II and gives the reason as "X-ray system C-arm casting may loosen and fall". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0584-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Essenta DR, a Stationary x-ray system, X-ray system Model: 712070 The Essenta DR is a multifunctional stationary X-ray system, in which the swivel arm rotation, the vertical movement of the swivel arm and the SID adjustment are motorized. It is designed for radiographic examination of the recumbent, standing or seated patient and also for lateral exposures. It enables the operator to quickly and safely move the unit to all the pre-programmed basic positions at the press of a button, using the autopositioning feature. Fine positioning of the tube/collimator and detector on the patient is easy to carry out using the command.
Product code
MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
Reason for recall
X-ray system C-arm casting may loosen and fall
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Philips Medical Systems issued a Urgent Field Safety Notice-Warning Label and Inspection, dated November 21, 2013, advising of potential of C-arm detaching and requesting a daily inspection and a safety warning label . If you need any further information or support concerning this issue, please contact your local Philips representative. 1-800-722-9377.
Quantity in commerce
424 units
Distribution
Worldwide Distribution - USA (Nationwide) including Puerto Rico and the countries of : Afghanistan, Argentina, Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hungary, India, Indonesia Iran,Italy, Jordan, Kazakstan, Kenya, Liberia, Lithuania, Morocco, Netherlands, New Zealand, Nigeria, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, South Korea, Spain, Sweden, Thailand, Turkey, Ukraine, United Arab, United Kingdom and Viet nam.
Product codes and lots
All Essenta DR x-ray Systems

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0584-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0584-2014Class IITerminatedVoluntary: Firm initiated