Siemens Artis Q and Artis Q biplane systems
Siemens Medical Solutions USA, Inc began a recall of Siemens Artis Q and Artis Q biplane systems Product Usage: Interventional Fluoroscopic X-ray on . FDA classified it as Class II and gives the reason as "There exists the possibility that during clinical operation, a breakdown of the x-ray tube could result in loss of x-ray imaging for the Artis Q and Artis Q biplane systems with a flat detector as 40 HDR (large 30x40)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1202-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Artis Q and Artis Q biplane systems Product Usage: Interventional Fluoroscopic X-ray
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- There exists the possibility that during clinical operation, a breakdown of the x-ray tube could result in loss of x-ray imaging for the Artis Q and Artis Q biplane systems with a flat detector as 40 HDR (large 30x40).
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens sent an Important Customer Safety Notice letter dated November 22, 2013, to affected customers. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to discontinue use until the affected product has been successfully replaced by a Siemen's representative.at no cost.
- Quantity in commerce
- 19
- Distribution
- USA Nationwide Distribution in the states of TN, MA, CA, IL, OH, FL, UT, WA, GA, IA, NY, MO, IN, and CO.
- Product codes and lots
- model numbers 10848280, 10848281, 10848282, and 10848283.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1202-2014 | Class II | Terminated | Voluntary: Firm initiated |