Z-0621-2014Class II

Integra Small Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311...

Integra LifeSciences Corporation / initiated 2013-11-13 / Terminated / root cause: packaging

Integra LifeSciences Corp. began a recall of Integra Small Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Made in the USA... on . FDA classified it as Class II and gives the reason as "Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type package were found prior to use but after the sterile package was open". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0621-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra Small Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536 USA Made in the USA NL 850-1121 Provides access to the lateral cerebral ventricles via hypodermic puncture. It is useful in obtaining CSF (cerebral spinal fluid_ samples for cytological and chemical studies, for monitoring ventricular fluid pressure and for ventricular drainage. The reservoir provides easy access to the lateral ventricles and to cystic tumors for the injections of chemotherapeutic agents and/or radio-isotopes. The rickham reservoir may be utilized in hydrocephalic patients as a component in systems designed to shunt CSF from the lateral ventricles into either the right atrium of the heart of the peritoneum.
Product code
JXG Shunt, Central Nervous System And Components
Reason for recall
Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type package were found prior to use but after the sterile package was open
Root cause tag
packaging
FDA root cause
Packaging
Action
Integra sent an Urgent Voluntary Medical Device Recall letter dated November 13, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their inventory for the affected product and if located stop using immediately. Customers were asked to complete the attached form and return it to Integra as indicated and keep a copy of the form for their records. Once the form was received by Integra they will send an RMA number, directions to return the product, and input an order to replace the quantity they indicated on the form. Customers with questions were instructed to call 1-855-532-1723. For questions regarding this recall call 855-532-1723.
Quantity in commerce
194 packages
Distribution
Worldwide Distribution - USA including FL, NE, WI, WA, CT, ND, AL, NY, NH, OR, OK, MN, and TX. Internationally to Belgium, Czech Republic, Germany, France, Great Britian, Italy, Poland, and Portugal.
Product codes and lots
1091271, 1094671, 1104165, 1104631, 1104813, 1110404, 1120638, 1120816, 1122316, 1123940, 11311659, 1132989, 1133427

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0621-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0621-2014Class IITerminatedVoluntary: Firm initiated
Z-0622-2014Class IITerminatedVoluntary: Firm initiated