Siemens ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400 Product Usage: The ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400 (also known as... on . FDA classified it as Class II and gives the reason as "The ADVIA Chemistry software systems using certain software versions allows the system to automatically run assay calibrations and quality control (QC)". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0940-2014
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400 Product Usage: The ADVIA Chemistry Systems models 1200, 1650, 1800 and 2400 (also known as Clinical Chemistry Systems) are high throughput instruments which perform assays for general and specialty chemistries, methods for specific protein measurement, drugs-of-abuse testing and therapeutic drug monitoring.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- The ADVIA Chemistry software systems using certain software versions allows the system to automatically run assay calibrations and quality control (QC).
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent an Urgent Medical Device Correct letter, dated November 6, 2013, to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to retain the recall notification letter with their laboratory records and forward to those who may have received the affected product.
- Quantity in commerce
- 3759
- Distribution
- Worldwide Distribution and USA Nationwide.
- Product codes and lots
- ADVIA 1200 V2.00 - 10386841 ADVIA 1200 V2.01 - 10469445 ADVIA 1650 V3.52 - 10318424 ADVIA 1650 V4.01 - 10285281 ADVIA 1800 V2.01 - 10639244 ADVIA 2400 V4.01 - 10639265