Z-0800-2014Class II

Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run...

Siemens Healthcare Diagnostics Inc. / initiated 2013-12-17 / Terminated / root cause: software

Siemens Healthcare Diagnostics, Inc. began a recall of Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine on . FDA classified it as Class II and gives the reason as "An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0800-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.
Root cause tag
software
FDA root cause
Software design
Action
The distribution of the Urgent Medical Device Correction letter scheduled to begin on December 17, 2013. x Field service personnel will be notified of the Urgent Medical Device Correction Letter via the Siemens Knowledge Base, as well as information regarding placing labels on the customer system to remind the customer that ///// is not a reportable result. x The Customers were instructed to contact Siemens Healthcare Diagnostics Technical Solution Center if they have technical questions regarding this issue. x Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction.
Quantity in commerce
1563
Distribution
Worldwide Distribution. USA nationwide and the country of Canada.
Product codes and lots
ADVIA 1200 with software versions: V2.00 and V2.01  ADVIA 1650 with software versions V3.52 and V4.01