Z-0605-2014Class II

VERIFY Dual Species Self-Contained Biological Indicator, Distributed by STERIS Corporation, Mentor, OH...

STERIS Corporation / initiated 2013-09-10 / Terminated / root cause: labeling and UDI

Steris Corporation began a recall of VERIFY Dual Species Self-Contained Biological Indicator, Distributed by STERIS Corporation, Mentor, OH. Intended for use in installation testing and routine... on . FDA classified it as Class II and gives the reason as "STERIS has identified that the population of the G. stearothermophilus spores that monitor STEAM sterilization cycles does not meet specifications for Lot # 131004 of the Verify Dual Species Self-Contained Biological...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0605-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
STERIS Corporation
Product
VERIFY Dual Species Self-Contained Biological Indicator, Distributed by STERIS Corporation, Mentor, OH. Intended for use in installation testing and routine monitoring of steam sterilization and ethylene oxide sterilization processes.
Product code
FRC Indicator, Biological Sterilization Process
Reason for recall
STERIS has identified that the population of the G. stearothermophilus spores that monitor STEAM sterilization cycles does not meet specifications for Lot # 131004 of the Verify Dual Species Self-Contained Biological Indicators. As a result the spore count on the affected product is lower than the labeled minimum and the product may incorrectly indicate that a 5-log spore kill was achieved.
Root cause tag
labeling and UDI
FDA root cause
Use error
Action
On September 10, 2013, the firm sent Urgent Voluntary Recall Notices to their customers.
Quantity in commerce
1,238 boxes
Distribution
Worldwide Distribution-USA (nationwide) including the states of AK, Al, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY, and the countries of BAHRAIN, CANADA, COLUMBIA, GERMANY, ITALY, MALAYSIA, PHILIPPINES, ROMANIA, SPAIN, and THE NETHERLANDS.
Product codes and lots
Lot #131004 Model #'s: S3060, S3061, S3065, S3069, LCB006 & LCB007.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0605-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0605-2014Class IITerminatedVoluntary: Firm initiated