Z-0713-2014Class II

pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units...

Radiometer America, Inc / initiated 2013-06-10 / Terminated / root cause: manufacturing process

Radiometer America Inc began a recall of pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro... on . FDA classified it as Class II and gives the reason as "RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0713-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Radiometer America, Inc
Product
pO2 D999 membrane units. Distributed by Radiometer America, Inc., Westlake, OH 44145 pO2 membrane units, ordering number 942-042 Intended use is in vitro testing of samples of whole blood for pO2 (partial pressure of oxygen).
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
RADIOMETER became aware that the regarded lot pO2 units 942-042R0971 were produced with inner solution for pCO2 membrane units. This solution is a salt solution similar to the pO2 inner solution, but also contains glycerol which can be reduced at the cathode causing a suddenly rising zero current, causing the membrane unit to fail the following Cal2 calibration.
Root cause tag
manufacturing process
FDA root cause
Mixed-up of materials/components
Action
Radiometer sent a Product Notification letter dated June 18, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory and remove any affected product, discard all affected product in their possession, and complete page 2 of the letter and return to their Radiometer representative. Radiometer will replace any quantity of affected product they have at their location. Customers with questions were instructed to contact Radiometer America Technicial Support at 1-800-736-0600, opt 4. For questions regarding this recall call 440-871-8900, ext 209.
Quantity in commerce
209 boxes/6 membranes
Distribution
Worldwide Distribution - US (nationwide) including states of: CO, CT, GA, MA, MO & MS and country of: Canada.
Product codes and lots
Part #: 942-042, Lot #: R0971

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0713-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0713-2014Class IITerminatedVoluntary: Firm initiated