Z-0570-2014Class II

Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the...

Philips North America LLC / initiated 2013-12-10 / Terminated / root cause: software

Philips Medical Systems, Inc. began a recall of Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and... on . FDA classified it as Class II and gives the reason as "MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0570-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician. The HeartStart MRx EtCO2 Option monitors carbon dioxide (CO2) and measures end-tidal carbon dioxide (EtCO2). The option provides an EtCO2 value and a CO2 waveform on the devices display.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg
Root cause tag
software
FDA root cause
Software design
Action
Philips Healthcare issued an Customer Information letter dated February 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer with the addendum to the Instructions for Use. If you have a ? in the parameter block and no CO2 waveform on the display, the waveform source is invalid. If invalid, check patient, confirm airway status and examine the FilterLine to make sure it is connected to the device and not kinked or pinched. For more information on the EtCO2 Option see Monitoring Carbon Dioxide chapter in the HeartStart MRx Instructions for Use. Should you have any questions or concerns, please contact your local Philips representative at 1-800-722-9377.
Quantity in commerce
18,529 units
Distribution
US Nationwide and OUS.
Product codes and lots
Software versions: F.02, R.02, 9.xx or any earlier versions   Addendum Models: M3532A, M3535A, M3536AM, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 M3535A- HeartStart MRx Defibrillator Monitor M3536A- HeartStart MRx Defibrillator Monitor EtC02 Options A04 A05 A06 All M3532A Upgrade

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0570-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0570-2014Class IITerminatedVoluntary: Firm initiated