Z-0593-2014Class II

Philips HeartStart MRx Monitor/Defibrillator Models: M3536A Options A20 - A27, HeartStart MRx ALS Monitor...

Philips North America LLC / initiated 2013-06-07 / Terminated / root cause: labeling and UDI

Philips Medical Systems, Inc. began a recall of Philips HeartStart MRx Monitor/Defibrillator Models: M3536A Options A20 - A27, HeartStart MRx ALS Monitor (Grey) The HeartStart MRx is for use for the... on . FDA classified it as Class II and gives the reason as "Error in the printing of the therapy energy setting labels on Philips HeartStart M3536A MRx Monitor/Defibrillators. The first energy setting is labeled 10 when it should be labeled1-10". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0593-2014
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips HeartStart MRx Monitor/Defibrillator Models: M3536A Options A20 - A27, HeartStart MRx ALS Monitor (Grey) The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Error in the printing of the therapy energy setting labels on Philips HeartStart M3536A MRx Monitor/Defibrillators. The first energy setting is labeled 10 when it should be labeled1-10
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
Philips sent a Customer Information Letter dated June 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. During an EI of the firm, the firm had not reported the correction to FDA. The Customer Information letter informs the customers what the behavior is and under what circumstances. The correction consisted of a replacement label that was provided free of charge to all units affected by this issue. Philips Healthcare representatives have been contact customers with affected devices to arrange for installation of the replacement label. To correct this issue, Philips will replace the label free of charge. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the replacement label. This issue does not affect current product shipments. I f you need any further information or support concerning this issue, please contact your local Philips representative at 1-800-722-9377. Philips apologizes for any inconvenience caused by this problem.
Quantity in commerce
93 units
Distribution
Worldwide Distribution - US Distribution including the states of IN, MS, TX and VA., and the countries of DENMARK, FRANCE, GERMANY, SPAIN and SWITZERLAND.
Product codes and lots
Serial numbers: US00558442 - US00558461 US00561519 US00558531 - US00558532 US00561578 US00558615 US00562454 US00558758 - US00558759 US00562504 - US00562527 US00560165 - US00560166 US00562640 - US00562641 US00560738 - US00560739 US00562832 US00560740 - US00560745 US00563125 US00560872 US00563231 - US00563243 US00560909 US00564032 US00561068 US00564177 US00561209 US00564409 US00561447 - US00561452 US00564793

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0593-2014
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0593-2014Class IITerminatedVoluntary: Firm initiated